ICU Medical CADD-Solis

Alarm: Reservoir volume empty

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the Reservoir volume empty alarm on an ICU Medical CADD-Solis infusion pump. The manual lists this as a high-priority alarm. It indicates the programmed reservoir volume has reached empty or the pump believes there is no remaining volume available for infusion.

This alarm can interrupt medication delivery and may require immediate clinical review, especially if the medication is time-sensitive, continuous, or critical to patient care.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Alarm Message

Check the Actual Reservoir or Medication Container

Verify the Programmed Reservoir Volume

Replace or Refill the Reservoir

Restart and Monitor the Infusion

If the Problem Persists

If the reservoir has medication remaining, the programmed reservoir volume has been verified, the reservoir or cassette has been replaced or refilled correctly, and the alarm still returns, external causes have been ruled out.

Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. Do not return the pump to patient use until the reservoir volume function, programming behavior, alarm response, and infusion operation have been verified.

Clinical Use Tip

Never troubleshoot an infusion pump on an active patient when medication delivery could be interrupted or changed unexpectedly. Move the patient to a backup pump or approved alternate delivery method first, then troubleshoot the pump when it is safe to do so.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis infusion pump displaying high-priority alarm: Reservoir volume empty."

Cause

What was observed during troubleshooting.

Example:
"Reservoir volume reached empty. Actual medication volume and programmed reservoir volume were reviewed with clinical staff. No pump damage, leak, or external setup issue found."

Resolution

What action was taken.

Example:
"Reservoir replaced or refilled per facility procedure. Pump programming confirmed before restart. Alarm cleared and pump operated normally during functional check."

Helpful Details to Include (If Known)

Final Thought

A Reservoir volume empty alarm should be handled as a clinical delivery issue first and an equipment issue second. Confirm patient safety, verify the actual reservoir status, confirm the programmed volume, and only return the pump to service after the alarm clears and operation is verified.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide