ICU Medical CADD-Solis

Alarm: Upstream occlusion

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

CADD-Solis

What This Guide Helps With

This guide helps troubleshoot the Upstream occlusion alarm on the ICU Medical CADD-Solis infusion pump. The manual lists upstream occlusion as a high-priority alarm. This condition means the pump detected, or is configured to detect, a restriction in the fluid path before the pump.

An upstream occlusion can interrupt medication delivery, delay therapy, or indicate a problem with the reservoir, cassette, tubing, clamps, or fluid path. The upstream occlusion sensor can be turned on or off, and the manual notes it is automatically disabled when medication cassette reservoirs are used.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Alarm Message

Check the Reservoir

Inspect the Cassette and Connection Area

Check Upstream Tubing and Clamps

Check for Upstream Fluid Path Restrictions

Confirm Alarm Behavior After External Corrections

If the Problem Persists

If the Upstream occlusion alarm continues after the reservoir, cassette, upstream tubing, clamps, and fluid path have been checked or corrected, external causes have been ruled out as much as practical.

Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. Do not return the pump to clinical use until the alarm condition has been evaluated and normal operation is confirmed.

Clinical Use Tip

Never troubleshoot an infusion pump alarm on an active patient when medication delivery may be unsafe. Move the patient to a backup pump or alternate approved delivery method first, then troubleshoot the CADD-Solis pump after patient care is protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ICU Medical CADD-Solis pump reported Upstream occlusion alarm during infusion setup and would not continue normal operation."

Cause

What was observed during troubleshooting.

Example:
"Found upstream clamp closed on administration set before pump. Reservoir, cassette, and tubing inspected for restriction."

Resolution

What action was taken.

Example:
"Opened clamp, corrected tubing path, verified disposable setup, and confirmed Upstream occlusion alarm cleared during functional check. Pump returned to service."

Helpful Details to Include (If Known)

Final Thought

An upstream occlusion alarm should be handled with patient safety first, then simple external checks before repair escalation. Most causes involve the reservoir, cassette, tubing, clamps, or fluid path before the pump. If those checks do not resolve the alarm, remove the pump from service and document the complaint, cause, and resolution clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide