ICU Medical Colleague

Air-in-Line False Alarms

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Colleague

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting ICU Medical Colleague infusion pumps that trigger air-in-line (AIL) alarms despite a properly primed and free-flowing line. These false alarms are often caused by setup issues, sensor contamination, or tubing irregularities rather than actual air in the fluid path. Proper troubleshooting ensures patient safety and prevents unnecessary pump replacement or service calls.

Step-by-Step Troubleshooting

Verify Alarm Message and Behavior

Inspect Tubing and Connections

Check for Air in Fluid Path

Inspect the Cassette and Sensor Area

Test with a Known-Good Cassette

Observe Environmental Factors

If the Problem Persists

After ruling out tubing, cassette, priming, and environmental causes, the false AIL alarms may be due to internal sensor or pump electronics issues.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump keeps alarming air-in-line even though line is primed and flowing normally.”

Cause

What was observed during troubleshooting.

Example:
“Observed tubing properly primed but AIL sensor contaminated with fluid residue; cassette slightly misaligned.”

Resolution

What action was taken.

Example:
“Cleaned sensor area, reseated cassette, replaced tubing if needed, confirmed alarm cleared. Pump returned to service.”

Helpful Details to Include

Final Thought

False air-in-line alarms can interrupt therapy and cause unnecessary alarm fatigue. Using a logical, stepwise approach focusing on tubing, cassettes, and sensor integrity ensures patient safety, avoids unnecessary repairs, and provides clear documentation for future reference.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide