ICU Medical Colleague

Occlusion Alarms with Free-Flowing Line

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Colleague

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting the ICU Medical Colleague infusion pump when it reports occlusion alarms despite a free-flowing line. These alarms often indicate resistance in the fluid path, but in many cases, they are caused by setup issues, tubing kinks, sensor misalignment, or cassette/door seating problems. The focus is on external, verifiable causes before assuming internal sensor or pump mechanism failure.

Step-by-Step Troubleshooting

Verify the Alarm Message and Context

Inspect Tubing Path and Connections

Check Cassette and Door Seating

Verify Infusion Setup

Swap Tubing or Cassette (If Available)

Check Pump Settings and Software

If the Problem Persists

After confirming tubing integrity, cassette seating, and correct setup, the issue may be internal (occlusion sensor, pump mechanism).

Action: Remove the pump from service, label it Out of Service, and send for repair or bench evaluation.

Escalating appropriately protects patient safety and prevents unnecessary troubleshooting of internal components on-site.

Clinical Use Tip

Do not troubleshoot on an active patient. Move the patient to another pump first.

Troubleshoot only when it is safe and will not interrupt therapy.

Ensure alarms are silenced only after patient safety is secured.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
Pump triggers occlusion alarm even though line is free-flowing.

Cause

What was observed during troubleshooting.

Example:
Tubing verified clear; cassette reseated; alarm persists. Likely internal sensor or mechanism issue.

Resolution

What action was taken.

Example:
Pump removed from service, labeled Out of Service, sent to Biomedical Engineering for repair.

Helpful Details to Include

Final Thought

Patient safety is the priority. False occlusion alarms can often be resolved by verifying tubing, cassette, and setup, but knowing when to escalate to repair is essential. Proper CCR documentation ensures clarity and continuity in maintenance records.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide