ICU Medical Plum 360 Infusion Pump

Air-in-Line Alarms

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Plum 360

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting air-in-line alarms on an ICU Medical Plum 360 infusion pump. These alarms occur when the pump detects air in the fluid path, which could be due to air bubbles in the syringe/line, improper priming, or loose connections. Early recognition and correction are critical to prevent air embolism risks to patients.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Check the Infusion Bag or Syringe

Inspect the Tubing and Connections

Prime the Line Correctly

Examine the Pump Air Sensor

Power Cycle the Pump

Reconnect and Test

If the Problem Persists

If the alarm continues after proper priming, tubing replacement, and inspection:

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump repeatedly alarms “Air-in-Line” despite infusion in progress.”

Cause

What was observed during troubleshooting.

Example:
“Air bubbles detected in the tubing; connections loose; fluid path not properly primed.”

Resolution

What action was taken.

Example:
“Line primed properly, tubing replaced, connections secured; alarm cleared. If unresolved, device removed from service.”

Helpful Details to Include

Final Thought

Clinical Engineering troubleshooting emphasizes patient safety, logical inspection of external causes, and knowing when to escalate to repair. Proper documentation using CCR ensures clarity and accountability for every intervention.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide