ICU Medical Plum 360

Alarm: Distal OCCLUSION! Check IV line and site.

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Plum 360

What This Guide Helps With

Distal OCCLUSION alarm causing stopped infusion due to downstream blockage, pressure buildup, or flow restriction in the IV line or patient site.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check IV Line for Kinks or Obstructions

Inspect Clamps and Stopcocks

Assess IV Site Condition

Check for Blocked or Occluded Accessories

Restart Infusion or Clear Alarm

Isolate Device vs. Setup

If the Problem Persists

If all external causes have been ruled out and the alarm continues:

Persistent occlusion alarms may indicate internal pressure sensing or flow control issues.

Clinical Use Tip

Never troubleshoot occlusion alarms on an active patient. Always transition the patient to a backup infusion device and confirm IV site patency before restarting therapy.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump alarming 'Distal OCCLUSION! Check IV line and site,' infusion stopped during use."

Cause

What was observed during troubleshooting.

Example:
"Kinked IV tubing downstream of pump causing flow restriction and pressure buildup."

Resolution

What action was taken.

Example:
"Straightened tubing, verified clamps open, restarted infusion. Alarm cleared and pump returned to service."

Helpful Details to Include (If Known)

Final Thought

Distal occlusion alarms are most commonly caused by simple external restrictions. Following a structured, external-first troubleshooting process ensures rapid resolution, maintains patient safety, and prevents unnecessary equipment downtime.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide