What This Guide Helps With

Distal OCCLUSION alarm during backprime or cassette check causing flow blockage and preventing proper priming or infusion.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Distal IV Line for Kinks or Obstructions

4. Inspect Patient-Side Components

5. Verify Cassette and Door Operation

6. Perform Backprime After Resolving Occlusion

7. Substitute IV Set if Needed

8. Return Device to Service if Resolved

If the Problem Persists

If all external causes have been ruled out and the alarm continues:

Clinical Use Tip

Never troubleshoot an occlusion on an active patient without confirming line patency. Pressure buildup can release as an unintended bolus.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump alarming 'Distal OCCLUSION! Resolve then Backprime' during backprime."

Cause

What was observed during troubleshooting.

Example:
"Kinked distal tubing near patient connection causing flow obstruction."

Resolution

What action was taken.

Example:
"Tubing repositioned, occlusion cleared, backprime successful, device tested."

Helpful Details to Include (If Known)

Final Thought

Most distal occlusion alarms are caused by simple downstream restrictions. A systematic external check ensures safe resolution and avoids unnecessary equipment removal.

That is successful troubleshooting.