This guide addresses the No Action Alarm for Line B, which occurs when no user interaction is detected after stopping Line B. This condition prevents infusion from continuing and may delay or interrupt therapy if not addressed.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Assess whether Line B infusion is required for active therapy
- If therapy is interrupted, transfer the patient to a backup infusion device
- Confirm patient continues receiving prescribed medication
Verify the Reported Alarm and Behavior
- Confirm the alarm message: “No Action Alarm! Start or Clear Line B.”
- Check if Line B was previously stopped or paused
- Determine if the pump has been idle for approximately 2 minutes
- Verify Line B is currently not infusing
Acknowledge the Alarm
- Press any key to acknowledge the alarm
- Observe if the alarm clears and allows interaction
Determine Intended Action for Line B
- If infusion should continue, press START for Line B
- If therapy is complete, clear Line B
- Ensure the correct state is selected based on clinical need
Check for Workflow or Setup Issues
- Verify staff are completing actions after stopping Line B
- Check for accidental stopping of Line B
- Ensure pump is not left idle during setup or transitions
Confirm Normal Operation
- Verify Line B is infusing if intended
- Ensure no additional alarms are present
- Monitor briefly to confirm stable operation
If the Problem Persists
If the alarm continues to occur unexpectedly:
- External and workflow-related causes have been ruled out
- Repeated alarms may indicate keypad, interface, or software issues
- Remove the device from service
- Label the device Out of Service
- Send for repair or bench evaluation
Clinical Use Tip
Never troubleshoot on an active patient. Always transfer the patient to a backup infusion device before addressing alarms or making adjustments.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"No Action Alarm! Start or Clear Line B displayed; Line B not infusing and alarm repeating after inactivity"
Cause
What was observed during troubleshooting.
Example:
"No operator action taken after stopping Line B; pump timed out awaiting user input"
Resolution
What action was taken.
Example:
"Acknowledged alarm, restarted Line B infusion (or cleared line as appropriate). Verified normal operation and no further alarms"
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
This alarm is a workflow-driven condition rather than a device failure. Recognizing inactivity alarms and responding appropriately ensures continuous therapy and prevents unnecessary escalation.
That is successful troubleshooting.