ICU Medical Plum 360

Alarm: OCCLUSION A! Check Line A.

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Plum 360

What This Guide Helps With

OCCLUSION A alarm indicating a blockage or air condition in proximal Line A, preventing proper fluid flow and stopping or preventing infusion delivery.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Line A for Kinks or Closed Clamps

Inspect Fluid Source and Container Setup

Check for Air in Line A

Verify Cassette and Door Seating

Restart Line A

If the Problem Persists

If the alarm continues after all external checks:

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot occlusions or air conditions while connected to a patient. Always transition the patient to a backup device before making adjustments.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump alarming 'OCCLUSION A! Check Line A.' and infusion stopped."

Cause

What was observed during troubleshooting.

Example:
"Kink in proximal tubing restricting flow into Line A."

Resolution

What action was taken.

Example:
"Removed kink, verified clamps open, cleared alarm, and confirmed normal infusion."

Helpful Details to Include (If Known)

Final Thought

Proximal occlusion alarms are often caused by simple external flow restrictions, but must be addressed immediately to ensure safe therapy. Always verify the full fluid path before escalating and document clearly.

That is successful troubleshooting.