ICU Medical Plum 360

Alarm: Proximal AIR Line A! Backprime.

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Plum 360

What This Guide Helps With

This guide addresses the “Proximal AIR Line A! Backprime.” alarm, which indicates that air has been detected in the proximal tubing. This condition stops infusion and requires removal of air before therapy can safely continue.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Inspect Proximal IV Line for Air (Most Common Cause)

4. Check for Loose Connections or Improper Setup

5. Perform Backprime Procedure

6. Reinstall Cassette and Restart Infusion

7. Substitute Administration Set if Issue Persists

If the Problem Persists

If the alarm continues after removing air, securing connections, and replacing the set, external causes have been ruled out. Internal air sensor or detection system failure is possible.

Clinical Use Tip

Never troubleshoot air-in-line alarms while connected to an active patient. Always disconnect and ensure the line is fully air-free before resuming infusion.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump alarming ‘Proximal AIR Line A! Backprime,’ infusion stopped."

Cause

What was observed during troubleshooting.

Example:
"Air present in proximal IV tubing due to incomplete priming or loose upstream connection."

Resolution

What action was taken.

Example:
"Removed air via backprime, secured connections, replaced tubing if needed, and verified proper operation."

Helpful Details to Include (If Known)

Final Thought

Air detection alarms must always be treated as a true safety concern. Confirm the source, remove all air, and only return the pump to service once safe operation is verified.

That is successful troubleshooting.