ICU Medical Plum 360

Alarm: Proximal OCCLUSION B! Check Line B.

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Plum 360

What This Guide Helps With

This guide addresses the “Proximal OCCLUSION B! Check Line B.” alarm, which indicates an upstream blockage in Line B. This condition stops infusion and prevents proper medication delivery.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Line B for Kinks or Closed Clamps

4. Check Fluid Source

5. Inspect Cassette and Pump Interface

6. Attempt to Restart Line B

7. Swap Administration Set

8. Isolate Device vs Setup

If the Problem Persists

If all external causes have been ruled out and the alarm continues:

Clinical Use Tip

Never troubleshoot occlusion alarms on an active patient. Always move the patient to a backup infusion device first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump alarming 'Proximal OCCLUSION B! Check Line B' during infusion; Line B stopped."

Cause

What was observed during troubleshooting.

Example:
"Kinked Line B tubing upstream of pump causing flow restriction."

Resolution

What action was taken.

Example:
"Straightened tubing and reopened clamp; verified infusion resumed without alarms."

Helpful Details to Include (If Known)

Final Thought

Proximal occlusion alarms are most often caused by simple upstream restrictions. Start with tubing, clamps, and fluid source before escalating. Knowing when to remove a device from service is critical for patient safety.

That is successful troubleshooting.