ICU Medical Plum 360

Alarm: Pump too high above patient. Lower pump or replace set.

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Plum 360

What This Guide Helps With

This guide addresses the alarm indicating the pump is positioned too high relative to the patient, causing increased backpressure and potential distal occlusion. This condition can stop infusion or prevent proper fluid delivery.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Pump Height Relative to Patient (Most Common Cause)

Lower the Pump to Correct Position

Inspect IV Tubing and Setup

Evaluate IV Set Condition

If the Problem Persists

If the alarm continues after correcting height, tubing, and IV set:

Clinical Use Tip

Never troubleshoot positioning-related alarms on an active patient. Always move the patient to a backup infusion device before making adjustments.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump alarming 'Pump too high above patient. Lower pump or replace set,' infusion not running."

Cause

What was observed during troubleshooting.

Example:
"Pump mounted above patient level causing backpressure and distal occlusion condition."

Resolution

What action was taken.

Example:
"Lowered pump to heart level, restarted infusion, and verified normal operation without recurring alarms."

Helpful Details to Include (If Known)

Final Thought

Position-related alarms are often simple but critical. Correct physical setup ensures accurate infusion delivery and prevents unnecessary escalation. Always verify external setup before assuming internal failure.

That is successful troubleshooting.