What This Guide Helps With
VTBI Completed Line B alarm indicating programmed volume finished, stopping infusion on Line B and requiring user action to continue or clear.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Assess the patient and confirm therapy status
- Transition to a backup infusion method if continuous therapy is required
- Verify ordered therapy is not interrupted
Verify the Reported Alarm and Behavior
- Confirm alarm message: “VTBI Completed Line B! Add more VTBI or Clear B.”
- Check if Line B infusion has stopped
- Determine if the alarm occurred at expected completion
- Check for any additional alarms
Confirm VTBI Settings and Infusion Completion
- Review programmed VTBI for Line B
- Verify VTBI has reached zero
- Check delivered volume against programmed amount
- Confirm therapy order matches delivered volume
Determine Next Clinical Action
- If continuing therapy: enter new VTBI and restart infusion
- If therapy is complete: stop Line B and clear the program
- Open cassette door if needed to reset programming
Check for Unintended Early Completion
- Verify correct VTBI was originally programmed
- Check infusion rate settings
- Ensure no incorrect configuration caused rapid delivery
Return Device to Service
- Confirm normal pump operation after action is taken
- Ensure no recurring alarms are present
- Verify infusion resumes correctly if restarted
If the Problem Persists
If the alarm occurs repeatedly without VTBI being reached or behaves inconsistently:
- External and programming causes have been ruled out
- This may indicate a sensing or software issue
- Remove the device from service
- Label the device Out of Service
- Send for repair or bench evaluation
Clinical Use Tip
Never troubleshoot on an active patient. Move the patient to a backup device before adjusting VTBI or restarting infusion.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Pump alarming 'VTBI Completed Line B! Add more VTBI or Clear B,' stopping infusion."
Cause
What was observed during troubleshooting.
Example:
"VTBI for Line B reached programmed limit, indicating normal infusion completion."
Resolution
What action was taken.
Example:
"Verified VTBI completion, entered new VTBI per clinical instruction, and confirmed normal operation."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Not all alarms indicate failure—some reflect normal therapy completion. Accurate interpretation prevents unnecessary troubleshooting while ensuring safe continuation of care.
That is successful troubleshooting.