ICU Medical Plum 360 Infusion Pump

Occlusion Alarms / Flow Blockages

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Plum 360

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting occlusion or flow blockage alarms on an ICU Medical Plum 360 infusion pump. These alarms occur when the pump detects high pressure in the infusion line, which may indicate a kink, obstruction, or improper setup. Most issues are external and do not require internal repair.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Check the Infusion Set and Tubing

Verify Syringe or Bag Placement

Check Flow Path and Pressure Limits

Examine Pump Settings

Power Cycle the Pump

Test with a Known Good Set

If the Problem Persists

All external causes have been ruled out (tubing, syringe, bag placement, flow path).

The pump’s pressure sensor or internal mechanism may be faulty.

Action:

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
User reports repeated occlusion alarms during infusion of normal saline.

Cause

What was observed during troubleshooting.

Example:
Tubing was kinked near the pump outlet; syringe initially misaligned.

Resolution

What action was taken.

Example:
Straightened tubing, reseated syringe, and verified free-flow. Pump now operates without occlusion alarms.

Helpful Details to Include

Final Thought

Logical, stepwise troubleshooting preserves patient safety and prevents unnecessary device downtime. External causes account for most occlusion alarms, so start there before assuming internal failure. Proper documentation using the CCR method ensures clarity for future reference and repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide