ICU Medical Plum 360 Infusion Pump

Pump Calibration or Accuracy Issues

On this page

Asset Type

Infusion Pump

Manufacturer

ICU Medical

Model

Plum 360

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting an ICU Medical Plum 360 infusion pump that is suspected of delivering inaccurate infusion volumes or rates. Symptoms may include:

This guide focuses on external and verifiable checks before assuming an internal calibration issue.

Step-by-Step Troubleshooting

Verify Setup and Programming

Check for Kinks or Obstructions

Inspect Cassette Placement and Condition

Check Power Source and Battery Status

Perform Manufacturer-Recommended Calibration Check

Use a Test Fluid Run (Optional, if safe)

If the Problem Persists

If the pump continues to deliver inaccurate volumes after all external checks:

Knowing when to stop prevents unsafe patient exposure and further device damage.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump delivering infusion rates inconsistent with programmed settings.”

Cause

What was observed during troubleshooting.

Example:
“External inspection revealed proper cassette placement, no tubing kinks, and correct programming. Self-test indicated potential calibration deviation.”

Resolution

What action was taken.

Example:
“Pump removed from service, labeled Out of Service, and sent for calibration verification by vendor. Patient transferred to verified pump.”

Helpful Details to Include

Final Thought

Accurate infusion delivery is critical for patient safety. Always start with external checks and documented programming verification before assuming internal calibration failure. Escalating appropriately ensures patient safety and maintains equipment reliability.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide