On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots weak or missing irrigation caused by fluid supply, tubing, cartridge installation, clamps, kinks, connections, settings, or external flow obstruction.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Surgical Continuity
If irrigation becomes unreliable during use, stop relying on the affected CUSA Clarity setup and provide an approved alternate method of irrigation or replacement system as directed by the surgical team.
Do not continue tissue fragmentation with an irrigation problem that could compromise safe operation.
Expected outcome: Patient care remains supported while the irrigation system is evaluated away from clinical dependence.
2. Confirm the Irrigation Symptom
Determine whether flow is completely absent, reduced, intermittent, delayed, or only affected at certain control settings.
Observe the system during a controlled nonclinical irrigation test.
Expected outcome: The exact irrigation behavior is reproduced and can be compared after each corrective action.
3. Verify Irrigation Fluid Supply
Confirm the correct irrigation source is present, contains adequate fluid, is appropriately positioned, and is not empty, disconnected, or inadvertently closed.
Inspect the fluid container and accessible supply connection.
Expected outcome: A usable irrigation fluid source is present and properly connected. If restoring the supply corrects flow, proceed to final verification.
4. Inspect the Entire External Irrigation Tubing Path
Trace the tubing from the fluid source to the handpiece. Look for clamps left closed, kinks, compression, pinching under equipment, disconnections, incorrect routing, or visible blockage.
Straighten or reconnect tubing using approved handling practices.
Expected outcome: The tubing path is continuous, unobstructed, and correctly routed.
5. Verify Cartridge or Tubing Installation
Confirm the irrigation cartridge or tubing segment is correctly installed in its designated external mechanism and fully seated.
Do not force the cartridge or bypass interlocks.
Expected outcome: The irrigation components are properly installed. If correcting installation restores steady flow, complete functional verification and stop troubleshooting.
6. Check Handpiece-Side Connections
Inspect the distal irrigation connection and tubing near the handpiece for incomplete seating, leakage, kinking, or damage.
Confirm the flue and related disposable setup are correctly assembled where relevant to the fluid path.
Expected outcome: Irrigation reaches the handpiece through intact and correctly connected tubing.
7. Review Irrigation Controls and Settings
Verify the selected irrigation setting is appropriate for the intended functional test and that staff have not inadvertently changed a control or preset.
Do not alter protected service-level configuration.
Expected outcome: Irrigation controls respond normally and commanded flow changes appropriately.
8. Substitute Known-Good Irrigation Consumables
If the tubing or cartridge remains suspect, replace the applicable disposable set with a known-good compatible set.
Repeat the irrigation test.
Expected outcome: Flow returns to normal with the replacement consumable, identifying the original set as the cause, or the problem persists.
9. Perform Final Functional Verification
With the complete setup assembled, verify continuous irrigation through the expected operating range used for functional testing and confirm there are no leaks, interruptions, or tubing movement.
Also verify handpiece recognition and overall system readiness.
Expected outcome: Irrigation is stable and repeatable. Troubleshooting can stop when required return-to-service checks pass.
10. Escalate Persistent Irrigation Failure
If fluid supply, tubing, cartridge installation, connections, disposables, and settings are correct but irrigation remains unreliable, stop external troubleshooting.
Expected outcome: The console or affected flow-control subsystem is referred for qualified service evaluation.
If the Problem Persists
Common external causes have been ruled out. The remaining issue may involve the irrigation drive mechanism, sensing, internal fluid-control hardware, control electronics, or another service-level condition.
The device should be:
- Removed from service.
- Labeled Out of Service.
- Sent for repair or bench evaluation.
- Evaluated using appropriate Integra LifeSciences documentation and approved test equipment.
- Repaired or configured only by qualified personnel.
- Functionally verified for stable irrigation before return to service.
Knowing when to stop after the complete external fluid path has been verified is proper troubleshooting.
Clinical Use Tip
Trace the irrigation path from the fluid source all the way to the handpiece rather than assuming the console is responsible for low flow.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported no irrigation flow from the CUSA Clarity handpiece during setup."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the irrigation tubing was kinked where it passed through the equipment setup."
Resolution
What action was taken.
Example:
"Corrected the tubing routing, verified continuous irrigation and normal handpiece function, and returned the system to service."
Helpful Details to Include (If Known)
- Whether flow was low, absent, or intermittent.
- Fluid source status.
- Tubing routing and clamp position.
- Cartridge installation.
- Handpiece-side connection condition.
- Visible leaks.
- Control setting observed.
- Results with replacement tubing or cartridge.
- Flow behavior before and after correction.
- Final system status.
Final Thought
Patient safety depends on reliable irrigation. Verify the simple fluid source and tubing path first, confirm installation and controls, use known-good consumables when appropriate, and escalate any persistent system-level failure.
That is successful troubleshooting.