Invacare Perfecto2

Low Oxygen Purity Alarm

On this page

Asset Type

Oxygen Concentrator

Manufacturer

Invacare

Model

Perfecto2

What This Guide Helps With

Troubleshooting low oxygen purity alarms caused by airflow restriction, filter issues, cannula problems, environmental conditions, or internal concentrator failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm whether the oxygen concentrator is actively supporting a patient.

If the patient requires continuous oxygen therapy, have clinical staff move the patient to another verified oxygen source before troubleshooting.

Expected outcome: The patient is not dependent on a concentrator with questionable oxygen output.

If patient safety is restored, continue troubleshooting only in a safe area.

Confirm the Reported Alarm

Power the Invacare Perfecto2 on and allow it to run long enough to confirm the reported low oxygen purity alarm.

Observe the alarm behavior, indicator lights, audible alarm, and whether the alarm occurs immediately or after warm-up.

Expected outcome: The issue is confirmed before unnecessary parts replacement or repair escalation.

Check the Oxygen Flow Setting

Verify the flowmeter is set within the prescribed and manufacturer-supported operating range.

Confirm the flow ball is stable and not set above the device’s rated output.

Expected outcome: Excessive flow demand is ruled out as a cause of low oxygen concentration.

If reducing the flow to an approved range clears the alarm and oxygen purity stabilizes, document the finding and stop.

Inspect the Cannula, Tubing, and Humidifier Bottle

Check for kinked tubing, crushed cannula lines, blocked nasal cannula tips, excessive tubing length, or a clogged humidifier bottle.

If a humidifier bottle is installed, verify it is not overfilled, cracked, cross-threaded, or creating back pressure.

Expected outcome: External patient accessory restrictions are corrected.

If replacing the tubing, cannula, or humidifier bottle resolves the alarm, return the unit only after performance is verified.

Check the Outlet Port and Oxygen Connection

Inspect the oxygen outlet fitting for obstruction, damage, looseness, or residue.

Confirm the tubing is fully seated and not leaking at the outlet connection.

Expected outcome: Oxygen flow is leaving the concentrator without obvious restriction or leakage.

Inspect the External Cabinet Filter

Check the external cabinet filter for dust, lint, pet hair, blockage, or moisture.

Clean or replace the filter according to facility procedure.

Expected outcome: Adequate room air intake is restored.

If the alarm clears after filter service and oxygen output verifies within acceptable limits, document the correction and stop.

Check the Room Environment

Verify the concentrator has adequate clearance around the cabinet and is not placed against a wall, curtain, bed linen, or furniture.

Confirm the unit is not operating in an excessively hot, dusty, or poorly ventilated area.

Expected outcome: The concentrator has enough airflow for cooling and sieve bed operation.

Why it matters: Restricted intake or poor cooling can reduce concentrator performance and trigger purity alarms.

Listen for Abnormal Compressor Operation

With the device running, listen for unusual compressor sounds, weak operation, loud rattling, rapid cycling, or air leaks.

Do not open the unit for internal diagnosis during basic troubleshooting.

Expected outcome: Obvious signs of internal compressor, valve, or pneumatic failure are identified for repair escalation.

Check for Obvious External Damage

Inspect the cabinet, power cord, caster area, outlet fitting, and filter cover for damage from impact, transport, liquid exposure, or contamination.

Expected outcome: Physical damage is identified before the unit is returned to service.

If damage is found, remove the concentrator from service.

Verify Output With Appropriate Test Equipment

In a Biomed-controlled setting, connect an oxygen analyzer and verify oxygen concentration after the unit has warmed up.

Confirm the analyzer is calibrated or verified before relying on the reading.

Expected outcome: The low oxygen purity complaint is either confirmed or ruled out using an independent measurement.

If oxygen concentration remains low after external checks, stop troubleshooting and escalate for repair.

If the Problem Persists

If the low oxygen purity alarm continues after checking flow setting, tubing, humidifier bottle, filters, airflow clearance, outlet connection, and oxygen output verification, common external causes have been ruled out.

The issue is likely internal, such as compressor weakness, sieve bed degradation, valve failure, internal leak, sensor issue, or control problem.

The device should be:

Knowing when to stop is proper troubleshooting. A concentrator that cannot reliably produce acceptable oxygen concentration should not remain available for patient use.

Clinical Use Tip

Do not troubleshoot a low oxygen purity alarm while the patient is relying on that concentrator. Move the patient to another verified oxygen source first, then evaluate the device in a safe, non-patient-use setting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported the Invacare Perfecto2 was alarming for low oxygen purity during patient oxygen setup."

Cause

What was observed during troubleshooting.

Example:
"External inspection found a heavily clogged cabinet filter and restricted airflow around the concentrator."

Resolution

What action was taken.

Example:
"Cleaned/replaced the cabinet filter, restored proper clearance, verified stable flow and acceptable oxygen concentration, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

Low oxygen purity alarms should be handled with patient safety first and a logical troubleshooting path. Rule out external airflow restrictions, accessory problems, excessive flow demand, and environmental causes before suspecting internal failure. If oxygen concentration cannot be verified as acceptable, remove the unit from service and document the complaint, cause, and resolution clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide