On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting no oxygen flow or low outlet flow caused by power, tubing, cannula, humidifier, filter, or external restriction issues.
Step-by-Step Troubleshooting
Ensure Patient Safety First
If the concentrator is in patient use and no flow or low flow is reported, have clinical staff place the patient on an alternate oxygen source immediately.
Expected outcome: The patient is supported by a known-good oxygen source before troubleshooting begins.
If alternate oxygen is needed, stop troubleshooting at the bedside until the patient is safe.
Verify the Reported Flow Problem
Turn the Invacare Perfecto2 on and allow it to run briefly in a safe, non-patient-care area.
Check whether the issue is:
- No flow at the outlet
- Low flow compared with the flowmeter setting
- Flowmeter ball not rising
- Flow present at the unit but not at the cannula
- Intermittent flow
Expected outcome: The actual symptom is confirmed before parts or accessories are replaced.
Check Power and Startup Condition
Confirm the unit powers on normally and does not shut down, alarm, or run unusually weak.
Verify the power cord is fully seated, the outlet is functional, and the concentrator is not connected through a questionable extension cord or power strip.
Expected outcome: The concentrator starts and runs with normal motor sound and indicator behavior.
If the unit will not power on or immediately alarms, troubleshoot the power/startup issue separately.
Check the Flowmeter Setting
Confirm the flow control knob is not fully closed or set below the requested flow rate.
Slowly adjust the flowmeter while observing the flowmeter ball.
Expected outcome: The flowmeter ball rises smoothly and can be adjusted to the prescribed flow range.
If the ball is stuck, erratic, or does not respond, the issue may involve the flowmeter or internal pneumatic path.
Remove Patient Tubing and Test Flow at the Concentrator Outlet
Disconnect the cannula, extension tubing, and humidifier bottle if installed.
Check for flow directly at the oxygen outlet using an appropriate flow test method.
Expected outcome: Flow is present at the concentrator outlet.
If flow returns after removing accessories, the concentrator may be functional and the restriction is likely external.
Inspect Cannula and Oxygen Tubing
Check the cannula and tubing for kinks, crushing, moisture buildup, clogged prongs, or excessive tubing length.
Swap with known-good oxygen tubing and cannula.
Expected outcome: Flow improves with known-good tubing.
If this resolves the issue, document the tubing or cannula restriction and return the concentrator to service only after confirming proper output.
Check the Humidifier Bottle, If Used
Inspect the humidifier bottle for cross-threading, blocked diffuser, overfilling, cracked cap, loose seal, or incorrect assembly.
Remove the humidifier bottle and connect tubing directly to the concentrator outlet for testing.
Expected outcome: Flow improves when the humidifier is bypassed if the bottle was causing restriction or leakage.
If the humidifier is defective, replace it and retest.
Inspect External Filters and Air Intake Area
Check the cabinet air filter and intake area for dust, lint, blockage, or placement too close to walls, bedding, curtains, or furniture.
Clean or replace external filters according to facility procedure.
Expected outcome: Air intake is clear and the unit can breathe properly.
A restricted intake can reduce concentrator performance and may contribute to low flow or alarms.
Check for Outlet or Fitting Damage
Inspect the oxygen outlet fitting for looseness, cracks, broken threads, missing adapter, or signs of damage from tubing being pulled.
Expected outcome: The outlet is secure and not leaking.
If the outlet fitting is damaged, remove the device from service for repair.
Listen for Abnormal Sounds or Leaks
With the unit running, listen for hissing, weak compressor operation, cycling irregularity, rattling, or air escaping from inside the cabinet.
Do not open the device for internal diagnosis during basic field troubleshooting.
Expected outcome: The concentrator runs smoothly without obvious leak sounds or unusual mechanical noise.
If abnormal sounds are present, internal pneumatic or compressor evaluation is likely needed.
Verify Output With a Flow Analyzer or Calibrated Flowmeter
Connect appropriate test equipment and compare actual output flow to the flowmeter setting.
Expected outcome: Delivered flow matches the selected setting within acceptable facility limits.
If the indicated setting and measured output do not match, remove the device from service for bench evaluation.
Check for Heat, Smell, or Signs of Damage
Inspect the cabinet, cord, plug, vents, and surrounding area for unusual heat, odor, discoloration, liquid intrusion, or physical damage.
Expected outcome: No unsafe condition is present.
If heat, smell, electrical damage, or liquid intrusion is found, stop testing and label the unit Out of Service.
If the Problem Persists
If no external restriction, accessory issue, power problem, filter blockage, or obvious user setup issue is found, the cause is likely internal.
Remove the Invacare Perfecto2 from service, label it Out of Service, and send it for repair or bench evaluation. Possible internal concerns may include the compressor, sieve bed performance, internal tubing, valves, regulator, flowmeter, or pneumatic leakage.
Knowing when to stop is proper troubleshooting. Do not return an oxygen concentrator to patient use unless flow output and overall operation are verified.
Clinical Use Tip
Do not troubleshoot a low-flow oxygen concentrator while a patient is depending on it. Move the patient to a verified backup oxygen source first, then evaluate the concentrator in a controlled area.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported the Invacare Perfecto2 had little to no oxygen flow at the patient cannula."
Cause
What was observed during troubleshooting.
Example:
"Found the humidifier bottle cap cross-threaded, causing leakage and reduced delivered flow."
Resolution
What action was taken.
Example:
"Reinstalled a known-good humidifier bottle, verified outlet flow with test equipment, and returned the concentrator to service."
Helpful Details to Include (If Known)
- Outlet tested and verified
- Flowmeter setting tested
- Actual measured output flow
- Cannula and tubing swapped
- Humidifier bottle bypassed or replaced
- External filter condition
- Alarm behavior or indicator lights
- Unusual sounds, heat, or smell
- Oxygen outlet fitting condition
- Final device status: returned to service or removed from service
Final Thought
For an oxygen concentrator with no flow or low output flow, start with patient safety and external restrictions before suspecting internal failure. Tubing, cannulas, humidifier bottles, filters, and outlet fittings are common causes that can be checked quickly. If verified output remains low after external causes are ruled out, remove the unit from service and document the findings clearly.
That is successful troubleshooting.