Cardinal Health Kangaroo ePump

Battery / Power Issues

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo ePump

What This Guide Helps With

This guide addresses situations where the Kangaroo ePump feeding pump:

The focus is on external, easily verifiable causes first—such as battery condition, charger, and connections—before assuming internal battery or electronics failure.

Step-by-Step Troubleshooting

Verify External Power Source

Why: If the outlet is dead, the pump cannot charge or operate.

Inspect Power Cord and AC Adapter

Why: Damaged cords can prevent charging or intermittent power.

Check Battery Seating

Why: Poor contact can prevent the pump from drawing power.

Verify Battery Charge Level

Why: A deeply discharged battery may need extended charging.

Swap with Known Good Battery (if available)

Why: Confirms whether the battery or pump electronics are at fault.

Inspect for Visible Damage

Why: External damage can cause internal power issues or short circuits.

Perform Soft Reset (if recommended by manufacturer)

Why: Clears temporary errors that can prevent boot-up.

If the Problem Persists

External causes (power source, cord, battery seating) have been ruled out.

Likely internal battery failure, charging circuit, or mainboard issue.

Action:

Note: Knowing when to escalate is proper troubleshooting.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump will not turn on and shows battery empty.”

Cause

What was observed during troubleshooting.

Example:
“Battery seated correctly but failed to hold charge; AC adapter verified working.”

Resolution

What action was taken.

Example:
“Pump removed from service, labeled Out of Service, sent for repair/bench evaluation.”

Helpful Details to Include

Final Thought

Patient safety and logic-based troubleshooting are paramount. Start with external verifications before assuming internal failure. Accurate documentation ensures proper follow-up and helps track recurring issues. Escalation when necessary is part of responsible device management.

That is successful troubleshooting.