Kangaroo OMNI Feeding Pump

Air-in-Line Alarms

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide provides a structured Clinical Engineering approach when a Kangaroo Omni feeding pump is generating Air-in-Line alarms.

The focus is on simple, external, and easily verifiable causes—such as improper priming, loose connections, empty feeding bags, or cassette seating—before assuming internal air sensor failure. These steps prioritize patient safety and proper escalation.

Step-by-Step Troubleshooting

Ensure the Patient Is Safe

Verify the Feeding Bag Is Not Empty

Inspect the Tubing for Visible Air

Re-Prime the Feeding Set

Check Cassette Seating

Inspect All Luer and Tubing Connections

Evaluate Formula Type and Foam

Swap the Feeding Set (If Needed)

Test the Pump Without a Patient

If the Problem Persists

If air-in-line alarms continue after:

The issue may involve an internal air detection sensor or cassette pressure system.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot an air-in-line alarm on an actively feeding patient.

Pause feeding and ensure patient stability first.

Air alarms are a safety feature designed to prevent air delivery into the GI tract. Always clear air completely before resuming feeding.

Work Order Documentation (CCR Method)

Complaint

What was reported by the clinical staff.

Example:
Nursing reports pump alarming “Air in Line” repeatedly and feeding will not continue.

Cause

What was observed during troubleshooting.

Example:
Air observed in tubing near cassette. Feeding bag nearly empty. Foam present in formula.

or

No visible air. Alarm persisted after re-priming. Suspected internal air sensor issue.

Resolution

What action was taken.

Example:
Replaced feeding bag and fully re-primed tubing. Alarm cleared. Pump tested and returned to service.

or

Replaced feeding set and verified proper cassette seating. Alarm resolved.

or

Removed pump from service for further evaluation after repeated alarms with no visible air source.

Helpful Details to Include (If Known)

Final Thought

Air-in-line alarms are commonly caused by incomplete priming, loose connections, or empty feeding bags. A calm, visual inspection and systematic correction of simple causes will resolve most cases. When external causes are ruled out, proper escalation protects both the patient and the integrity of the equipment program. Clear documentation ensures traceability and supports future trend analysis.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide