Asset Type
Manufacturer
Model
What This Guide Helps With
Helps isolate abnormal video artifacts caused by camera cables, connectors, optics, video paths, displays, interference, or external component failure.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Reliable Visualization
Do not continue using an image with significant noise, lines, color distortion, or dropout when visualization could affect the procedure.
Move the clinical team to a verified alternate imaging path before disturbing cables or components.
Expected outcome: Reliable visualization is maintained during troubleshooting.
2. Characterize the Artifact
Determine whether the problem is fixed-pattern noise, horizontal or vertical lines, color shift, intermittent black frames, complete dropout, or a combination.
Note whether the artifact appears continuously, after warming up, or when a cable is moved.
Expected outcome: The artifact is described clearly enough to compare across components and displays.
If the image is stable and artifact-free, troubleshooting can stop after verification.
3. Determine Whether the Artifact Is in the Source or Display Path
Compare the image on another connected known-good display or output if available.
If the artifact is present on every display, focus on the camera and source path. If only one display is affected, focus on the downstream video path.
Expected outcome: The problem is narrowed to the source side or display side.
4. Inspect the Camera Cable and Connections
Inspect the camera cable, connector, and strain relief for damage, contamination, sharp bends, or looseness.
During off-patient testing, gently reposition the cable while observing the image.
Expected outcome: The image remains stable during normal cable positioning.
If the artifact changes when the cable moves, remove the affected camera head or cable assembly from service.
5. Inspect Video Output Cables and Routing
Inspect external video cables and any accessible routing connections between the controller and display.
Reseat connectors and bypass unnecessary adapters or routing equipment when possible.
Expected outcome: The video path is securely connected and does not introduce noise or dropout.
If bypassing the installed route removes the artifact, escalate the AV or infrastructure path.
6. Inspect the Optical Path
Check accessible optical surfaces for contamination, moisture, or damage that could be perceived as image shading or color distortion.
Verify the telescope and coupler are seated correctly.
Expected outcome: The optical path is clean and does not contribute visible artifacts.
If cleaning or reseating resolves the image problem, proceed to final verification.
7. Substitute Known-Good Components
When available, substitute a verified camera head, telescope, coupler, video cable, or display one item at a time.
Expected outcome: The artifact follows a specific component or is isolated to the controller or facility video path.
If a failed external component is identified, remove it from service.
8. Check for Environmental or Interference Factors
Observe whether the artifact appears only when nearby equipment is activated or when cables are routed near other powered devices.
Do not alter facility grounding or electrical systems as a troubleshooting shortcut.
Expected outcome: The video remains stable under normal room operating conditions.
If environmental interference is suspected, escalate to the appropriate technical group.
9. Perform Final Functional Verification
Operate the camera through normal modes and gently move accessible cables within normal use limits.
Verify that the image remains clear, color remains stable, and no dropout or repeating artifact occurs.
Expected outcome: Video remains consistently stable and suitable for clinical use.
If these checks pass, the system may be returned to service.
If the Problem Persists
Camera cables, video paths, displays, optics, environmental factors, and known-good substitutions have been checked.
Remove the affected equipment from service, label it Out of Service, and send it for repair or bench evaluation. Persistent artifacts may involve internal camera imaging hardware, signal processing, controller electronics, output hardware, firmware, or infrastructure.
Evaluate using appropriate Karl Storz documentation and approved test equipment. Avoid internal board-level troubleshooting unless specifically authorized and trained.
Return the equipment to service only after stable video has been verified under normal operating conditions.
Clinical Use Tip
Intermittent video dropout should be treated as a reliability failure even when the picture returns on its own.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported intermittent lines and brief video dropout when the camera cable was repositioned."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the artifact consistently followed movement at the camera-head cable strain relief."
Resolution
What action was taken.
Example:
"The affected camera head was removed from service, a verified replacement was installed, and stable artifact-free video was confirmed."
Helpful Details to Include (If Known)
- Type of image artifact
- When the artifact occurred
- Displays or outputs affected
- Camera-cable condition
- Whether movement changed the symptom
- Video cables and routing tested
- Optical condition
- Known-good substitutions
- Environmental correlation
- Final image status
Final Thought
Image artifacts are best isolated by following the signal from camera to display. Check external cables, optics, routing, and environment before assuming internal failure, and escalate any persistent reliability problem appropriately.
That is successful troubleshooting.