Kendall SCD 700 Series

Pressure Too Low or Inconsistent

On this page

Asset Type

Sequential Compression Device (SCD)

Manufacturer

Cardinal Health

Model

Kendall SCD 700 Series

What This Guide Helps With

This guide provides a structured Clinical Engineering approach when a Kendall SCD 700 Series delivers compression pressure that appears:

The focus is on logical, external, and easily verifiable causes before assuming internal pump, valve, or pressure sensor failure.

Step-by-Step Troubleshooting

1. Confirm Therapy Mode and Active Cycle

Incorrect mode selection or incomplete startup can mimic low-pressure delivery.

2. Check Sleeve Application and Fit

Oversized or loosely applied sleeves often feel like low pressure even when the unit is functioning properly.

3. Inspect Tubing for Leaks or Obstructions

Even a small air leak significantly reduces delivered compression pressure.

4. Reseat All Connections at the Ports

Partially seated connectors are a common cause of inconsistent pressure.

5. Swap Sleeves and Tubing

This step helps isolate accessory failure from internal device failure.

6. Listen to Pump Operation

Rapid cycling without building pressure often indicates a leak or internal valve malfunction.

7. Check for Alarms or Error Messages

8. Compare Both Channels

If the Problem Persists

If pressure remains low or inconsistent after completing these steps, common remaining causes include:

These are not field-correctable.

Next Steps:

Knowing when to stop bedside troubleshooting is proper Clinical Engineering practice.

Clinical Use Tip

Do not troubleshoot pressure concerns while sleeves are actively connected to a patient.

Transfer the patient to a functioning SCD unit first.

Verify DVT prophylaxis therapy continues uninterrupted.

Perform troubleshooting only when patient safety is ensured.

Never compromise active therapy to investigate equipment performance.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Compression pressure weak and inconsistent on left leg.”

Cause

What was observed during troubleshooting.

Example:
“Sleeves and tubing inspected and swapped; issue remained on left port. Suspected internal valve or pump degradation.”

or

“Issue followed sleeve when swapped. Sleeve found leaking.”

Resolution

What action was taken.

Example:
“Replaced defective sleeve; unit returned to service.”

“Removed device from service and sent for bench evaluation due to suspected internal pressure channel failure.”

Helpful Details to Include (If Known)

Final Thought

Most low-pressure complaints are caused by loose connections, accessory leaks, incorrect sleeve sizing, or incomplete setup — not internal pump failure.

A logical, step-by-step approach:

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide