Kendall SCD 700 Series

One or More Sleeves Not Inflating

On this page

Asset Type

Sequential Compression Device (SCD)

Manufacturer

Cardinal Health

Model

Kendall SCD 700 Series

What This Guide Helps With

This guide provides a structured Clinical Engineering approach when one or more sleeves connected to a Kendall SCD 700 Series unit are not inflating properly.

Common symptoms include:

This guide focuses on simple, external, and easily verifiable causes before assuming internal pump or valve failure.

Step-by-Step Troubleshooting

Confirm the Unit Is Powered and Therapy Is Started

A common oversight is the device being powered but not actively running therapy.

Check Sleeve and Tubing Connections

Loose or partially seated connectors are the most common cause of one sleeve not inflating.

Inspect the Tubing for Kinks or Damage

Even a small leak will prevent the system from building pressure.

Inspect the Sleeves

Visually inspect for:

Ensure sleeves are wrapped snugly but not overly tight. Improper fit can prevent pressure buildup or cause uneven compression.

Swap Sleeves to Isolate the Issue

If the issue follows the sleeve → the sleeve is likely defective.
If the issue stays with the same port → suspect an internal valve or channel issue.

This step is critical for isolating accessory failure versus device failure.

Check for Alarm Messages

Document any alarm messages exactly as displayed, such as:

Alarm wording helps determine whether the system is sensing a leak or failing to generate pressure.

Listen for Pump Operation

Start therapy and listen carefully:

Avoid opening the device. Internal repair is not a bedside procedure.

If the Problem Persists

If the sleeves still do not inflate after completing all external checks, you have ruled out:

The issue is likely internal, such as:

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot sleeve inflation while actively connected to a patient requiring therapy.

If compression therapy is clinically indicated:

Patient safety and uninterrupted therapy take priority over troubleshooting speed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:

“One sleeve not inflating; right leg receiving no compression.”

Cause

What was observed during troubleshooting.

Example:

“Tubing and sleeve inspected. Swapped sleeves; issue followed right sleeve. Sleeve leaking.”
or
“Swapped sleeves and ports. Issue remained on left port. Suspected internal valve failure.”

Resolution

What action was taken.

Example:

“Replaced defective sleeve; unit tested and returned to service.”
or
“Removed device from service and sent to shop for internal repair.”

Helpful Details to Include (If Known)

Final Thought

Most sleeve inflation problems are accessory-related rather than internal failures. Logical, external troubleshooting protects patients, prevents unnecessary equipment swaps, and supports accurate documentation. Escalate only after simple causes have been ruled out.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide