LivaNova S5

Battery Backup, Power Supply, or CAN-Bus Communication Problem

On this page

Asset Type

Heart-Lung Machine

Manufacturer

LivaNova

Model

S5

What This Guide Helps With

Troubleshoots backup-power, external power, and system communication problems using power-source checks, module isolation, connections, and safe functional verification.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain a Fully Functional Backup System

Do not troubleshoot unstable S5 power or system communication while a patient depends on the affected equipment. A loss of power distribution or shared communication can affect multiple perfusion functions simultaneously.

Follow the perfusion contingency plan and transition to verified backup support as required.

If there is smoke, odor, overheating, arcing, liquid intrusion, or repeated power loss, remove power when clinically safe and take the system out of service.

Expected outcome: Patient support is maintained independently of the suspect S5 power or communication system.

Once safely isolated from clinical use, troubleshooting may continue.

2. Confirm the Failure Pattern

Determine whether the problem involves:

Record all displayed messages and indicator states.

Expected outcome: The failure is categorized as source-power, backup-power, or shared communication related.

If the system remains stable through repeated controlled tests after a correctable external condition is found, troubleshooting can stop.

3. Verify Facility AC Power

Inspect the power cord, plug, strain relief, and accessible input connections. Confirm the receptacle is powered using approved test equipment or a known-good source.

Check whether the problem occurs at one outlet only or follows the S5.

Expected outcome: A stable facility AC source is confirmed.

If an outlet or external power problem is identified and correction restores normal operation, troubleshooting can stop after system verification.

4. Inspect External Power Connections

With the S5 removed from service, inspect all accessible power connections associated with the console and connected modules.

Look for:

Do not open internal power supplies.

Expected outcome: External power connections are fully seated and free of visible damage.

If correcting an external power connection restores reliable operation, troubleshooting can stop after testing.

5. Evaluate Battery-Supported Operation

Using approved procedures, determine whether the system:

Do not intentionally run backup batteries to exhaustion.

Expected outcome: Backup power is recognized and supports the system as intended during the approved test.

If a replaceable approved battery component is confirmed defective and replacement restores normal operation, troubleshooting can stop after verification.

6. Inspect External Communication Connections

Because shared CAN-bus communication problems may affect multiple modules, inspect accessible intermodule communication cables and connectors for:

Reseat only approved external communication connections.

Expected outcome: External communication connections are secure and undamaged.

If reseating a communication connection restores all affected modules and stability is confirmed, troubleshooting can stop.

7. Determine Whether One Module Is Disrupting the System

If permitted by approved service procedures, identify whether the communication problem occurs only when a specific external module or cable is connected. Use controlled isolation or known-good substitution without altering protected internal bus termination or configuration.

Do not remove modules required for active clinical support.

Expected outcome: The problem is isolated to a specific external module/cable or remains system-wide.

If a defective external module connection or cable is identified and correction restores stable communication, troubleshooting can stop after final verification.

8. Verify Normal Configuration and Module Population

Confirm that connected modules match the normal system configuration and that no component has been intentionally removed, disabled, or reconfigured.

Do not change restricted CAN-bus parameters, internal addressing, or termination.

Expected outcome: Installed modules and the authorized configuration are consistent.

If an authorized configuration correction resolves communication, troubleshooting can stop.

9. Perform Final System Verification

After correction, verify:

Complete all applicable return-to-service testing.

Expected outcome: The complete S5 system remains stable across approved power and communication checks.

If all checks pass, troubleshooting is complete.

10. Escalate Persistent Power or CAN-Bus Problems

If the system has repeated power loss, unexplained resets, battery-transfer problems, or persistent communication loss after external causes are ruled out, stop troubleshooting.

Do not open internal power supplies, repair bus electronics, change internal termination, or replace circuit boards without authorized service procedures.

Expected outcome: A system with unresolved shared power or communication risk remains unavailable for patient use.

If the Problem Persists

Facility power, external power connections, backup-power behavior, module cabling, and normal configuration have been evaluated. Remaining possibilities may involve internal power distribution, battery-management circuitry, internal CAN-bus components, module electronics, or protected configuration.

The S5 should be:

Before return to service, complete applicable electrical safety, backup-power, communication, module-recognition, alarm, and full-system functional testing.

Knowing when to stop external troubleshooting is especially important when one fault can affect multiple perfusion functions.

Clinical Use Tip

A shared power or communication problem can affect several S5 modules at once, so verify the complete heart-lung machine rather than only the first function that appeared to fail.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Perfusion reported that several S5 modules intermittently disappeared from the console during pre-case setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a shared external communication connector partially seated after the system had been moved."

Resolution

What action was taken.

Example:
"Clinical Engineering reseated and secured the connector and verified stable module communication, repeated startup, and normal system operation."

Helpful Details to Include (If Known)

Final Thought

Shared power and communication faults deserve conservative handling because they can affect multiple perfusion functions simultaneously. Verify facility power and external connections first, isolate modules logically, avoid unauthorized internal bus work, complete full-system verification, and document the event clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide