Masimo Rad-97

Low Perfusion Causes Persistent No-Read Condition

On this page

Asset Type

Pulse Oximeter

Manufacturer

Masimo

Model

Rad-97

What This Guide Helps With

Persistent no-read conditions associated with weak peripheral perfusion, poor sensor placement, motion, cable problems, or unsuitable measurement sites.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Monitoring

Do not troubleshoot unreliable monitoring equipment while a patient depends on it. Transfer the patient to another verified pulse oximeter or use an appropriate alternate monitoring method before removing or repositioning the sensor.

Confirm that clinical staff are assessing the patient independently and not relying on an absent saturation reading.

Expected outcome: Continuous patient monitoring is maintained while the Rad-97 is evaluated.

2. Confirm the Exact Reported Condition

Ask whether the no-read condition occurs continuously or intermittently, on one patient or multiple patients, and at a specific site or under particular conditions such as cold extremities, hypotension, movement, transport, or vasoconstriction.

Observe the displayed indicators, messages, Signal IQ information, and pleth behavior without changing clinical settings unnecessarily.

Expected outcome: The failure is confirmed and distinguished from an isolated patient-specific low-perfusion event.

3. Inspect the Measurement Site

Check whether the selected site is cold, poorly perfused, compressed, edematous, injured, positioned below a blood pressure cuff, or affected by excessive motion.

Have clinical staff select another clinically appropriate, better-perfused site when permitted.

Expected outcome: A stable pleth waveform and numerical reading appear after correcting the measurement site. If readings remain stable, troubleshooting can stop.

4. Verify Sensor Placement and Alignment

Inspect the sensor for correct positioning, full contact, proper emitter-to-detector alignment, and secure application without excessive pressure.

Confirm that reusable sensors are clean and that disposable sensors are not loose, folded, contaminated, or repeatedly reapplied beyond their intended condition.

Expected outcome: The sensor detects pulsatile flow and produces a consistent reading. If successful, troubleshooting can stop.

5. Reduce Motion and Environmental Interference

Stabilize the monitored extremity and cable. Check for tremor, shivering, patient movement, transport vibration, or cable tugging.

Shield the sensor from intense direct light when appropriate and verify that no external source is interfering with the optical path.

Expected outcome: Signal quality improves and the no-read condition clears. If stable readings are restored, troubleshooting can stop.

6. Inspect the Sensor, Patient Cable, and Connections

Examine the sensor, patient cable, adapter, and monitor connector for contamination, bent contacts, damaged insulation, strain, looseness, or incomplete insertion.

Disconnect and reconnect each accessible connection securely. Do not force connectors.

Expected outcome: All connections remain secure and the signal returns without manipulation. If so, troubleshooting can stop.

7. Substitute Known-Good Accessories

Test the monitor using a compatible known-good sensor and patient cable approved for the Rad-97.

Test the suspect sensor and cable on another compatible verified monitor when available.

Expected outcome: The substitution identifies whether the problem follows the sensor, cable, or monitor. Replace the failed external accessory and stop if normal performance is restored.

8. Compare Performance on a Suitable Test Subject or Simulator

After removing the device from patient use, test it on a suitable staff test subject or approved pulse oximetry simulator according to facility procedures.

Do not use a simulator result alone to dismiss a genuine patient perfusion concern.

Expected outcome: The monitor acquires and maintains a stable signal under controlled conditions.

9. Perform Final Functional Verification

Verify stable saturation, pulse rate, pleth waveform, signal quality indication, sensor-off response, and applicable alarms using approved test methods.

Confirm that the device functions without cable manipulation or intermittent signal loss.

Expected outcome: The Rad-97 performs consistently and is suitable for return to service.

10. Escalate an Unresolved No-Read Condition

If the Rad-97 repeatedly fails to acquire a signal with compatible known-good sensors, cables, suitable sites, and controlled testing, stop external troubleshooting.

Expected outcome: The device is removed from clinical use and routed for qualified service evaluation.

If the Problem Persists

Common external causes such as poor site selection, low peripheral perfusion, motion, sensor placement, cable condition, and accessory compatibility have been ruled out.

The remaining cause may involve the internal signal-processing path, connector assembly, configuration, software, or another service-level condition. Remove the device from service, label it Out of Service, and send it for repair or bench evaluation using current manufacturer documentation and approved test equipment.

Repairs or configuration changes must be performed only by qualified personnel. Complete functional and alarm verification before returning the device to service. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A no-read condition can reflect the patient’s physiology rather than equipment failure; maintain alternate monitoring and clinical assessment during device evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Rad-97 remained in a no-read condition on a patient with weak peripheral perfusion."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found poor sensor contact at a cold, poorly perfused finger and verified that the monitor operated normally with a known-good sensor at an alternate site."

Resolution

What action was taken.

Example:
"Repositioned the sensor at a suitable site, confirmed a stable pleth waveform and numerical readings, tested alarm response, and returned the device to service."

Helpful Details to Include (If Known)

Final Thought

Protect the patient first, verify site and accessory conditions before assuming internal failure, escalate persistent no-read conditions appropriately, and document the confirmed cause and final verification clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide