On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Intermittent sensor detection or readings caused by loose connections, damaged cables, worn adapters, strain, contamination, or monitor connector problems.
Step-by-Step Troubleshooting
1. Protect the Patient and Replace Unreliable Monitoring
Do not troubleshoot an intermittent monitoring cable while a patient depends on it. Transfer the patient to another verified monitor or replace the affected monitoring path immediately.
Expected outcome: Patient monitoring continues without intermittent signal loss.
2. Confirm the Intermittent Condition
Determine whether the signal drops during movement, cable flexing, bed adjustment, transport, or repositioning.
Observe messages, pleth waveform, Signal IQ, sensor status, and numerical readings without deliberately stressing the cable while it is connected to a patient.
Expected outcome: The reported intermittent failure is confirmed and associated with a specific condition.
3. Inspect the Sensor Connection
Check that the sensor is fully seated in the patient cable or adapter and that locking features engage correctly.
Inspect for contamination, moisture, bent contacts, worn fit, or damage.
Expected outcome: The sensor connection remains secure and readings stabilize. If successful, troubleshooting can stop after verification.
4. Inspect the Patient Cable and Adapter
Examine the full cable length, strain reliefs, connectors, and adapter housing for cuts, kinks, crushing, discoloration, exposed conductors, or prior repairs.
Remove any damaged cable or adapter from service.
Expected outcome: Only intact external accessories remain in use.
5. Inspect the Rad-97 Connector
Check the monitor connector for contamination, looseness, damaged contacts, or evidence that the plug does not remain fully seated.
Do not probe or reshape contacts.
Expected outcome: The monitor connector is clean, undamaged, and holds the cable securely.
6. Reconnect and Route the Cable Correctly
Disconnect and reconnect the cable without force. Route it to avoid tension, sharp bends, snagging, bed mechanisms, wheels, and patient movement.
Expected outcome: The signal remains stable during normal repositioning.
7. Substitute a Known-Good Patient Cable and Adapter
Use compatible approved known-good accessories while keeping the same monitor and sensor.
Expected outcome: The intermittent condition disappears, identifying the original cable or adapter as defective. Replace it and stop after final verification.
8. Test the Suspect Accessory on Another Compatible Monitor
When appropriate, test the suspect cable or adapter on another verified compatible device under controlled conditions.
Expected outcome: The problem follows the accessory or remains with the original Rad-97.
9. Test With a Known-Good Sensor
Use a compatible known-good sensor to ensure that an intermittent sensor lead is not being mistaken for a cable or adapter problem.
Expected outcome: The signal remains stable or the sensor is identified as the cause.
10. Perform Controlled Movement Testing
With the device removed from patient use, use an approved simulator or suitable test method and gently reposition the cable through normal expected movement.
Do not apply excessive bending or destructive stress.
Expected outcome: No signal dropout, sensor disconnect, or waveform interruption occurs.
11. Complete Functional and Alarm Verification
Confirm stable readings, pleth waveform, Signal IQ, sensor disconnect alarm, and recovery after reconnection.
Expected outcome: The monitoring path is reliable and suitable for return to service.
12. Escalate a Suspected Monitor Connector Failure
If the problem remains with multiple known-good sensors, cables, and adapters, remove the Rad-97 from service.
Expected outcome: The monitor is labeled Out of Service and routed for connector or internal input evaluation.
If the Problem Persists
External causes involving loose seating, cable damage, adapter wear, sensor defects, strain, and contamination have been ruled out.
The remaining cause may involve the monitor input connector, internal cable assembly, signal-processing path, or another service-level condition. Remove the device from service, label it Out of Service, and send it for repair or bench evaluation using current manufacturer documentation and approved test equipment.
Do not perform deep disassembly or connector repair without authorization. Complete signal, alarm, and intermittent-movement testing before return to service.
Clinical Use Tip
Route patient cables away from siderails, bed joints, wheels, and transport handles to prevent recurring strain and intermittent monitoring loss.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the Rad-97 repeatedly lost the pulse oximetry signal when the patient cable was moved."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found an intermittent open condition near the patient cable strain relief, confirmed by substitution with a known-good cable."
Resolution
What action was taken.
Example:
"Replaced the damaged patient cable, verified stable readings during controlled movement, tested sensor-disconnect alarms, and returned the device to service."
Helpful Details to Include (If Known)
- Condition that triggered signal loss
- Displayed sensor or cable message
- Pleth and Signal IQ behavior
- Sensor, cable, and adapter identification
- Visible damage or contamination
- Monitor connector condition
- Known-good substitution results
- Controlled movement test results
- Alarm verification
- Final device status
Final Thought
Protect the patient from intermittent monitoring, trace the signal path from sensor to monitor, replace defective external accessories before assuming internal failure, and verify reliability under normal movement.
That is successful troubleshooting.