Masimo Rad-97 Pulse Oximeter

Power / Battery Issues

On this page

Asset Type

Pulse Oximeter

Manufacturer

Masimo

Model

Rad-97

What This Guide Helps With

This guide addresses situations where the Masimo Rad-97 pulse oximeter:

The focus is on external, easily verifiable causes first—such as battery condition, charger, and connections—before assuming internal battery or board failure.

Step-by-Step Troubleshooting

Verify Battery Charge and Connections

Why it matters: Dirty or misaligned contacts can prevent proper charging, mimicking battery failure.

Test With a Known Good Power Source

Why it matters: Eliminates external power issues before assuming battery failure.

Remove and Reinstall the Battery

Why it matters: Sometimes battery reseating resolves poor contact or software detection errors.

Try a Known Good Battery

Why it matters: Confirms whether the original battery is at fault without opening the device.

Check for Firmware or Device Alerts

Why it matters: Internal battery management alerts can help determine if the issue is software or hardware related.

If the Problem Persists

External causes (power source, charging contacts, battery seating) have been ruled out.

Device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Stopping at this stage is proper troubleshooting and avoids patient risk.

Clinical Use Tip

Do not troubleshoot on an actively monitored patient. Move the patient to a backup pulse oximeter first.

Only perform troubleshooting in a safe, controlled environment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Rad-97 would not power on; patient monitoring interrupted.”

Cause

What was observed during troubleshooting.

Example:
“Battery failed to hold charge; charger and power source verified.”

Resolution

What action was taken.

Example:
“Device removed from service, labeled Out of Service, sent to Biomedical Engineering for battery replacement and functional testing.”

Helpful Details to Include

Final Thought

Power and battery issues are commonly caused by easily fixed external factors like charger, outlet, or battery contacts. Proper troubleshooting avoids unnecessary internal disassembly, protects patient safety, and ensures accurate documentation for follow-up repair.

That is successful troubleshooting.