Medivators Advantage Plus

Drain Failure or Chamber Will Not Empty

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Medivators

Model

Advantage Plus

What This Guide Helps With

Addresses retained chamber fluid, slow draining, or drain alarms caused by external hose restriction, facility drain blockage, positioning, or cycle interruption.

Step-by-Step Troubleshooting

1. Protect Patients and Reprocessing Workflow

Treat any endoscope from an interrupted or incomplete cycle as not fully reprocessed. Quarantine it and transfer reprocessing to another verified AER when available.

Do not open, move, or disconnect the AER while retained fluid could spill or expose personnel to chemicals.

Expected outcome: Staff, endoscopes, and the surrounding area are protected before troubleshooting begins.

2. Confirm the Exact Drain Condition

Record the exact displayed message, cycle phase, chamber involved, and whether fluid remains visibly present. Determine whether the chamber does not drain at all, drains slowly, or drains and then refills.

Check for unusual pump sounds, leakage, odor, or fluid around the base.

Expected outcome: The drain failure is accurately characterized and any immediate leak or chemical-exposure hazard is identified.

3. Check the Facility Drain for Backup

Inspect the accessible drain connection and surrounding area for standing water, slow drainage, or backup from sinks and nearby equipment. Ask whether Facilities or Environmental Services recently reported a blocked drain.

Do not continue cycling the AER into a drain that is backing up.

Expected outcome: The facility drain accepts flow normally. If a building drain obstruction is corrected and the chamber drains normally, troubleshooting can stop after verification.

4. Inspect the External Drain Hose

Inspect the drain hose from the AER to the facility drain. Look for:

Correct only accessible external routing problems.

Expected outcome: The drain hose is secure, leak-free, unobstructed, and routed to permit normal drainage.

5. Verify the AER Is Properly Positioned

Check that the unit has not shifted, settled unevenly, or been pushed against the wall in a way that pinches the drain hose. Verify that recent cleaning, construction, or service has not changed the drain connection.

Do not level or reposition the AER beyond facility-approved practices.

Expected outcome: The AER and external drain path are positioned without strain or restriction.

6. Check for Visible Chamber Obstruction

With the AER safe, idle, and handled according to approved procedures, inspect the chamber for visible debris, packaging material, loose labels, or incorrectly placed accessories that may obstruct accessible drain openings.

Do not remove fixed strainers, panels, or internal plumbing components unless the manufacturer procedure specifically authorizes it.

Expected outcome: No visible foreign material or incorrectly positioned accessory is blocking the accessible drainage path.

7. Verify Load and Accessory Placement

Confirm that the endoscope, tubing, connectors, lids, and accessories were placed according to the department’s validated process. Incorrect routing or an item resting over a drain area may interfere with fluid movement.

Correct the setup and keep the original load quarantined until a complete reprocessing cycle is performed.

Expected outcome: The chamber load and accessories do not interfere with drainage.

8. Perform a Safe Restart or Drain Attempt

When chemical exposure risk is controlled and approved procedures allow it, perform the manufacturer-authorized drain, cancel, or restart action. Do not repeatedly command the unit to drain if the fluid level does not change or leakage develops.

Observe the external drain hose and facility drain during the attempt.

Expected outcome: Fluid exits steadily without leakage or backup. If the chamber empties normally and the alarm does not return, proceed to verification.

9. Perform a Controlled Functional Verification

Run the approved empty-cycle or verification process. Confirm that:

Expected outcome: The complete verification process passes. The issue is resolved, and troubleshooting can stop.

10. Stop and Escalate When Drainage Remains Unreliable

If the external drain, hose routing, chamber condition, and load setup are acceptable but the chamber still will not empty, remove the AER from service. Do not disassemble internal valves, pumps, sensors, or chemical plumbing.

Expected outcome: The AER is prevented from processing endoscopes until reliable drainage is restored.

If the Problem Persists

Common external drain restrictions and setup problems have been ruled out. The remaining cause may involve an internal drain valve, pump, level-sensing system, internal tubing restriction, control fault, or facility drain-capacity problem requiring measured service evaluation.

The AER should be:

Return to service requires successful drainage, leak inspection, alarm verification, and completion of the required cycle-performance testing. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Do not release an endoscope from any cycle that stopped with fluid remaining in the chamber.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported that the Advantage Plus stopped during the drain phase and left water in the chamber."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external drain hose compressed between the AER and the wall."

Resolution

What action was taken.

Example:
"The hose was repositioned and secured, the chamber was safely drained, and an approved verification cycle completed with normal draining and no leaks."

Helpful Details to Include (If Known)

Final Thought

Protect personnel and endoscopes first, verify the complete external drain path before assuming internal failure, confirm reliable emptying through a full test, escalate safely, and record the cause and outcome clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide