Medivators Advantage Plus

Leak Test Failure Preventing Cycle Initiation

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Medivators

Model

Advantage Plus

What This Guide Helps With

Leak test failures that prevent cycle start, typically caused by scope connection issues, leaks in the scope, or improper hookup.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: No risk of using a potentially contaminated or unverified scope

Verify the Reported Error and Behavior

Expected: Leak test consistently fails before disinfection begins

Why it matters: Confirms this is not a one-time or user setup issue

Check Scope Connection Integrity (Most Common Cause)

Expected: All connections are tight and correctly matched

Why it matters: Air leaks at connection points will fail the pressure-based leak test

Inspect the Endoscope for External Damage or Known Leaks

Expected: No visible damage or known leak issues

Why it matters: A damaged scope will fail the AER leak test regardless of setup

Verify Correct Leak Test Adapter Use (If Applicable)

Expected: Adapter is correct, intact, and properly seated

Why it matters: Faulty adapters commonly introduce leaks into the system

Check for Kinked or Blocked Tubing

Expected: Tubing is unobstructed and properly routed

Why it matters: Pressure inconsistencies can trigger leak test failure

Perform a Manual Leak Test (If Available Per Facility Practice)

Expected: Scope holds pressure with no visible leaks

Why it matters: Confirms whether the issue is the scope vs the AER

Retry Cycle with a Different Scope (Isolation Step)

Expected: Cycle proceeds normally with alternate scope

Why it matters: Helps isolate whether failure is scope-related or AER-related

If the Problem Persists

If all external causes (connections, adapters, tubing, and scope condition) have been ruled out, the issue is likely internal to the AER (pressure sensing, valves, or leak test system).

Remove the unit from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never rely on a scope that fails a leak test. Always remove it from service immediately and use a verified, properly functioning scope for patient care.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"AER fails leak test and will not start disinfection cycle."

Cause

What was observed during troubleshooting.

Example:
"Loose scope channel connection causing air leak during pressure test."

Resolution

What action was taken.

Example:
"Reconnected scope properly and verified successful cycle initiation."

Helpful Details to Include (If Known)

Final Thought

Leak test failures are most often caused by setup or scope-related issues, not the AER itself. A structured, external-first approach ensures patient safety, protects equipment, and avoids unnecessary service calls. Proper documentation supports traceability and quality assurance.

That is successful troubleshooting.