On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Addresses missing, incomplete, faded, or failed cycle records caused by paper, printer power, connections, queue, settings, or external reporting problems.
Step-by-Step Troubleshooting
1. Protect Traceability and Reprocessing Records
Do not discard or release required cycle documentation without following the facility’s approved downtime process. Confirm whether the AER itself shows a successfully completed cycle before addressing only the output problem.
Do not interpret a missing printout as proof that the cycle failed or succeeded.
Expected outcome: Endoscope traceability is preserved while record output is restored.
2. Confirm the Exact Output Failure
Determine whether the issue involves:
- No printout
- Blank paper
- Faded print
- Partial record
- Paper jam
- Repeated records
- Wrong cycle information
- A report that exists electronically but does not print
- Failure affecting one or all cycles
Record any displayed printer or export message.
Expected outcome: The problem is separated into printer hardware, media, queue, configuration, or data-generation categories.
3. Verify Cycle Completion and Stored Record
Check whether the completed cycle appears in the AER’s normal record or history view. Confirm the correct endoscope, operator, date, time, and cycle status.
Do not recreate or alter the clinical record merely to produce a printout.
Expected outcome: The source cycle record is confirmed as present and accurate, or a broader record-generation problem is identified.
4. Check Printer Power and Status
Verify that the printer is powered, online, and free of obvious fault indicators. Check its approved power source, power cord, and accessible connection to the AER.
If the printer uses a separate outlet, verify that the outlet supports another low-risk test device according to facility practice.
Expected outcome: The printer has stable power and is in a ready state.
5. Inspect Paper and Media Loading
Check that the correct approved paper or media is installed with the proper orientation. Inspect for:
- Empty roll or tray
- Incorrect paper type
- Reversed thermal paper
- Wrinkled or damp paper
- Misaligned roll
- Open cover
- Paper scraps
- Visible jam
Do not insert tools into the printer mechanism.
Expected outcome: Approved media is installed correctly and feeds freely. If proper loading restores printing, continue to verification.
6. Inspect External Printer Connections
Inspect and reseat accessible printer cables only while the devices are safely idle. Look for:
- Loose connectors
- Damaged cable insulation
- Bent pins
- Strain at the connector
- Liquid contamination
- Connection to the wrong port
Use a known-good approved cable when available.
Expected outcome: The printer connection is secure and undamaged. If a known-good cable restores output, troubleshooting can stop after verification.
7. Check Output Selection and Queue
Verify that the correct printer or output destination is selected through authorized user settings. Check for a paused printer, accumulated queue, or a record waiting for manual release.
Do not change protected network, system, or service configuration without authorization.
Expected outcome: The intended output destination is active and able to receive the cycle record.
8. Perform a Controlled Test Print or Reprint
Use an approved test-print, stored-record reprint, or nonclinical verification function. Confirm that the printout is complete, legible, and matches the selected record.
Avoid creating duplicate clinical documentation without marking or managing it according to policy.
Expected outcome: The printer produces a complete and accurate record. The issue is resolved, and troubleshooting can stop.
9. Verify the Full Cycle Record Workflow
Run the approved verification cycle and confirm:
- Successful cycle completion
- Electronic record creation
- Correct date and time
- Correct identifiers
- Complete printed or exported record
- No printer or communication alarm
Expected outcome: The complete documentation pathway functions correctly.
10. Stop and Escalate When Record Output Remains Unreliable
If paper, power, connections, queue, and output settings are correct but records remain missing or inaccurate, remove the printer function or AER from service as required by facility traceability policy.
Do not disassemble the printer or alter protected system configuration without approved documentation.
Expected outcome: Reprocessing is not continued without a reliable documentation method.
If the Problem Persists
Common power, media, connection, queue, and destination causes have been ruled out. The remaining issue may involve the printer mechanism, print interface, local software, data-generation process, networked report destination, or configuration requiring authorized service.
The device should be:
- Removed from service when required by traceability policy
- Labeled Out of Service
- Sent for repair or qualified evaluation
- Evaluated using appropriate manufacturer documentation and approved test equipment
- Repaired or configured only by qualified personnel
Return to service requires verification that complete and accurate cycle records are created and delivered through the required documentation pathway. Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Confirm the electronic cycle result before treating a missing printout as either a successful or failed reprocessing cycle.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterile Processing reported that completed Advantage Plus cycles were not producing printed records."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the external printer powered off after its power strip had been switched off during cleaning."
Resolution
What action was taken.
Example:
"Printer power was restored, a stored record printed correctly, and a verification cycle generated a complete and legible cycle report."
Helpful Details to Include (If Known)
- Exact printer message
- Whether cycle record exists electronically
- Printer power and status indicators
- Outlet tested
- Paper type and orientation
- Jam or media condition
- Cable condition
- Output destination
- Reprint result
- Record accuracy
- Verification-cycle result
- Final device status
Final Thought
Preserve traceability, verify the stored cycle record before assuming data loss, check simple printer and media causes first, validate the complete output pathway, escalate appropriately, and document the final record status.
That is successful troubleshooting.