Medivators Advantage Plus

Rinse Water Quality, Conductivity, or Final Rinse Alarm

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Medivators

Model

Advantage Plus

What This Guide Helps With

Addresses final-rinse or water-quality alarms caused by facility water changes, treatment-system problems, expired filters, contamination, flow instability, or sensing faults.

Step-by-Step Troubleshooting

1. Protect Patients and Reprocessing Workflow

Treat any endoscope from a cycle stopped by a final-rinse or water-quality alarm as incompletely reprocessed. Quarantine it and use another verified AER when available.

Do not bypass the alarm or release the endoscope based only on earlier completed phases.

Expected outcome: Endoscopes are not used after an unverified final rinse.

2. Confirm the Exact Alarm

Record the exact message and determine whether it refers to:

Ask whether the issue began after filter service, plumbing work, sanitation, or a water interruption.

Expected outcome: The alarm is accurately categorized before troubleshooting begins.

3. Check for Facility Water Events

Contact Facilities or review department notices for:

Compare with another AER or water-quality-dependent device on the same supply when appropriate.

Expected outcome: A facility-wide water condition is either identified or ruled out.

4. Inspect Incoming Water Appearance and Supply Condition

Where facility procedures permit, check the incoming water source for visible discoloration, sediment, odor, air, or unstable flow. Confirm the supply valve is fully open and the external hose is not kinked.

Do not independently sample or certify water quality without approved equipment and procedure.

Expected outcome: Incoming water appears normal and reaches the AER without restriction.

5. Review Filter and Water-Treatment Status

Check documented replacement status for all applicable external and AER-associated water filters. Inspect accessible housings for leaks, incorrect valve alignment, overdue labels, or recent incomplete service.

Verify that facility water-treatment equipment serving the AER is operational.

Do not bypass filters or treatment stages.

Expected outcome: Required filtration and treatment components are in service and within documented status.

6. Check for Recent Filter Replacement or Sanitation

Determine whether the alarm began immediately after filter replacement, sanitation, plumbing work, or prolonged shutdown. Verify that required flushing, priming, or post-service verification was completed by qualified personnel.

Do not reset water-quality warnings without confirming the underlying condition.

Expected outcome: Post-service air, residue, or incomplete flushing is identified and corrected through the approved process.

7. Verify Drain and Flow Conditions

Check that the chamber drains fully and incoming flow is stable. Retained fluid, poor drainage, or inconsistent filling can affect the final-rinse process and may appear as a water-quality alarm.

Inspect external supply and drain hoses for restriction.

Expected outcome: The AER fills, circulates, and drains normally without mixing or retention caused by an external restriction.

8. Check Environment and Chemical Storage

Confirm that detergent, disinfectant, and other approved consumables are correctly connected and not leaking into the wrong area. Inspect for spilled chemical near water connections or filter housings.

Do not attempt to neutralize or flush chemical contamination without an approved procedure.

Expected outcome: No visible chemical cross-contamination or incorrect consumable connection is present.

9. Perform Approved Water-System Verification

After external utilities, filters, treatment equipment, and post-service steps are confirmed, run the approved empty-cycle, rinse, or water-system verification process. Observe for:

Expected outcome: The AER accepts the rinse-water condition and completes the verification cycle. The issue is resolved, and troubleshooting can stop.

10. Stop and Escalate When Water Quality Cannot Be Verified

If the facility supply, filters, treatment system, flow, drainage, and consumable connections are acceptable but the alarm remains, remove the AER from service.

Do not adjust conductivity calibration, bypass sensors, or access internal water-quality components without approved manufacturer procedures.

Expected outcome: The AER is prevented from processing endoscopes until final-rinse quality can be reliably verified.

If the Problem Persists

Common facility water, filtration, treatment, flow, drainage, and post-service causes have been ruled out. The remaining issue may involve an internal conductivity or water-quality sensor, calibration condition, internal contamination, valve issue, control fault, or a facility water condition requiring laboratory or engineering evaluation.

The AER should be:

Return to service requires acceptable water-system verification, successful final-rinse operation, alarm clearance, leak inspection, and complete cycle-performance testing. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A completed wash phase does not compensate for a failed or unverified final rinse.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported repeated final-rinse water-quality alarms on the Advantage Plus after facility plumbing work."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering and Facilities found sediment loading in the external pretreatment filter following restoration of the water supply."

Resolution

What action was taken.

Example:
"Facilities replaced the approved pretreatment filter, the supply was flushed according to procedure, and the AER completed the required verification cycle without water-quality alarms."

Helpful Details to Include (If Known)

Final Thought

Protect the patient by treating every final-rinse alarm as an incomplete cycle, verify facility water and treatment components before assuming internal failure, complete approved verification after correction, escalate unresolved conditions, and document the full water-quality event.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide