Medivators Advantage Plus

Touchscreen Unresponsive or Software Freeze

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Medivators

Model

Advantage Plus

What This Guide Helps With

Addresses frozen screens, delayed touch response, or software lockups caused by contamination, power interruption, peripheral faults, environmental conditions, or software problems.

Step-by-Step Troubleshooting

1. Protect Patients and Reprocessing Workflow

If the touchscreen freezes during a cycle, treat the cycle as incomplete unless the AER clearly completes and records it according to the validated process. Quarantine the endoscope and move work to another verified AER.

Do not disconnect power while chemical movement, draining, or an active cycle may still be occurring unless an emergency requires it.

Expected outcome: The affected endoscope and staff remain protected while the equipment state is assessed.

2. Confirm the Exact Condition

Determine whether:

Record all displayed messages before attempting a restart.

Expected outcome: The failure is characterized as a touch-input, display, software, or broader power condition.

3. Check Whether a Cycle Is Active

Observe status indicators, pump sounds, fluid movement, alarms, and any connected printer or record output. Do not open the chamber or interrupt utilities solely because the display appears frozen.

Follow the approved emergency or interrupted-cycle procedure if the AER state cannot be determined.

Expected outcome: The unit’s operating state is identified without creating a chemical or reprocessing hazard.

4. Inspect and Clean the Touchscreen Surface

Check for:

Clean and dry the screen using an approved method.

Expected outcome: The touchscreen surface is clean, dry, undamaged, and free of physical interference. If touch response returns, continue to verification.

5. Check External Peripherals

Disconnect only approved removable peripherals that are not required for safe operation, such as an external barcode scanner or printer, after the unit is safely idle. A failed peripheral or cable can sometimes affect software response.

Do not disconnect network, chemical, water, or internal connections as a diagnostic shortcut.

Expected outcome: A faulty external peripheral is isolated, or the screen remains frozen with peripherals removed.

6. Verify Stable Power

Inspect the power cord, plug, strain relief, outlet, and any approved power-distribution device. Check for evidence of recent power interruption, loose plugs, tripped circuits, or facility electrical work.

Do not continue using an AER that loses power unpredictably.

Expected outcome: The AER has a secure and stable power source.

7. Perform an Authorized Controlled Restart

Once the cycle state is safe and the affected endoscope remains quarantined, use the normal shutdown or restart sequence. Avoid repeated forced power cycles.

Observe startup for error messages, unusual delays, or recurrence of the freeze.

Expected outcome: The AER starts normally and the touchscreen responds across the full display. If it freezes again, continue troubleshooting.

8. Check Environment and Ventilation

Verify that external vents are unobstructed and the room is not unusually hot, cold, or humid. Look for condensation, nearby steam sources, blocked airflow, or heat-producing equipment.

Do not open service panels to cool the unit.

Expected outcome: The AER operates in a stable environment with unobstructed ventilation.

9. Perform a Full Functional Verification

Test the touchscreen across normal authorized screens and run the approved empty or verification cycle. Confirm:

Expected outcome: The touchscreen and software remain stable through the complete test. The issue is resolved, and troubleshooting can stop.

10. Stop and Escalate When Freezing Recurs

If the touchscreen is clean, power is stable, peripherals are ruled out, and environmental conditions are acceptable but the freeze returns, remove the AER from service. Do not access internal computing hardware, reinstall software, or use unauthorized service tools.

Expected outcome: An AER with unreliable controls is prevented from processing endoscopes.

If the Problem Persists

Common screen contamination, peripheral, power, and environmental causes have been ruled out. The remaining issue may involve the touchscreen assembly, internal computer, storage device, application software, operating system, internal connection, or control electronics.

The AER should be:

Return to service requires reliable control response, successful startup, alarm verification, complete cycle testing, and confirmation that cycle records are accurate. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A frozen display during reprocessing requires the endoscope to remain quarantined until cycle completion can be positively verified.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported that the Advantage Plus touchscreen stopped responding at the cycle-selection screen."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found dried chemical residue along the lower edge of the touchscreen, causing continuous false touch input."

Resolution

What action was taken.

Example:
"The screen was cleaned and dried using the approved method, and the AER completed a full verification cycle with reliable touch response and no freeze."

Helpful Details to Include (If Known)

Final Thought

Protect the endoscope and determine the AER’s actual cycle state first, rule out screen contamination, peripherals, power, and environment before assuming internal failure, verify stable operation, escalate recurring freezes, and document the event precisely.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide