Medivators DSD-201 / DSD Edge Series

Cycle Not Starting or Aborting Mid-Cycle

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Medivators

Model

DSD-201 / DSD Edge Series

What This Guide Helps With

Cycle will not initiate or stops mid-process due to door, water, chemical, or connection issues affecting proper reprocessing.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: No potentially contaminated scope is used
Why it matters: Incomplete disinfection creates infection risk

Verify the Reported Issue

Expected: Clear identification of failure point
Why it matters: Narrows troubleshooting to a specific system function

Check Door Closure and Latch Status (Most Common Cause)

Expected: Door fully seals and is recognized by system
Why it matters: Safety interlocks prevent operation if sealing is compromised

Verify Water Supply and Pressure

Expected: Unit fills properly without delay or alarm
Why it matters: Water flow is required for all cleaning and rinsing stages

Check Chemical Levels and Connections

Expected: Proper chemical detection and dosing
Why it matters: AER will abort if required chemicals are unavailable

Inspect Scope Connections and Channel Hookups

Expected: Secure and correct scope connection
Why it matters: Leak test or flow failure will prevent or stop the cycle

Check for Leak Test Failures

Expected: Leak test passes successfully
Why it matters: System will not proceed if scope integrity is compromised

Inspect Drain and Waste Line

Expected: Waste drains freely without alarm
Why it matters: Inability to drain interrupts cycle progression

Power Cycle the Unit

Expected: System resets and cycle starts normally
Why it matters: Clears minor software or sensor faults

If the Problem Persists

If all external causes have been ruled out, the issue is likely internal (sensors, valves, pumps, or control system faults).

Remove the unit from service, label it Out of Service, and send for repair or further bench evaluation.

Knowing when to stop prevents unnecessary risk and avoids extended downtime from improper troubleshooting.

Clinical Use Tip

Never attempt to troubleshoot or restart a failed cycle for a scope intended for immediate patient use.

Always move the case to another properly functioning AER and ensure proper reprocessing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"AER cycle would not start and intermittently aborted during operation."

Cause

What was observed during troubleshooting.

Example:
"Door latch not fully engaging, preventing system from recognizing sealed condition."

Resolution

What action was taken.

Example:
"Adjusted door closure and verified proper latch engagement; unit tested and passed full cycle."

Helpful Details to Include (If Known)

Final Thought

Effective AER troubleshooting focuses on external systems first—door sealing, water, chemicals, and connections. These account for most failures. When those are ruled out, escalation is the correct and safe decision. Always prioritize infection control and document findings clearly to support compliance and patient safety.

That is successful troubleshooting.