Medtronic Capnostream 20p

Alarm: EtCO2 HIGH ALARM

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot an EtCO2 HIGH ALARM on a Medtronic Capnostream 20p capnography monitor. This alarm means the measured EtCO2 is above the upper alarm limit and may indicate a real patient ventilation issue, airway issue, rebreathing condition, incorrect alarm limit, sampling accessory problem, or unreliable EtCO2 measurement.

Because elevated EtCO2 can reflect patient compromise, this alarm should always be treated as a patient safety alarm first. Equipment troubleshooting should only begin after the patient has been assessed and safe monitoring is maintained.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Clinical Response Before Equipment Troubleshooting

Check the Alarm Limit Setting

Inspect the CO2 Sampling Line or Airway Adapter

Check the Capnogram Waveform

Check for Rebreathing or External Setup Issues

Compare With Another Monitor or Measurement Source if Needed

Check Power, Connections, and Physical Condition

Remove From Service if Equipment Fault Is Suspected

If the Problem Persists

If clinical causes, alarm settings, sampling accessories, power, external connections, and setup issues have been ruled out, the persistent alarm may indicate a device measurement issue or internal CO2 monitoring problem.

Remove the Capnostream 20p from service, label it Out of Service, and send it for repair or bench evaluation.

Knowing when to stop is proper troubleshooting. Do not keep troubleshooting on an active patient or continue using a monitor with unresolved EtCO2 alarm concerns.

Clinical Use Tip

Never troubleshoot this alarm on an active patient as if it is only an equipment issue. A high EtCO2 alarm can indicate a real ventilation problem. Move the patient to a backup monitor or approved alternate monitoring method first if the Capnostream 20p must be removed from patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p reported with EtCO2 HIGH ALARM. User stated EtCO2 value was above the upper alarm limit during patient monitoring."

Cause

What was observed during troubleshooting.

Example:
"Alarm confirmed. Clinical evaluation required first due to elevated EtCO2 reading. External checks included alarm limit review, sampling line inspection, FilterLine replacement, waveform observation, power check, and physical inspection."

Resolution

What action was taken.

Example:
"Verified alarm behavior and confirmed clinical staff response. Replaced questionable sampling accessory and checked external connections. If alarm cleared with correct setup and clinical staff approved, device returned to service. If alarm persisted with known-good accessories or questionable EtCO2 readings, device removed from service for bench evaluation."

Helpful Details to Include (If Known)

Final Thought

An EtCO2 HIGH ALARM must be treated as a patient condition first. Clinical Engineering should support safe troubleshooting by confirming the alarm, checking simple external causes, verifying accessories and alarm settings, and escalating the device if the reading appears unreliable. Clear documentation protects the patient, the staff, and the equipment history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide