On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide helps troubleshoot an EtCO2 HIGH ALARM on a Medtronic Capnostream 20p capnography monitor. This alarm means the measured EtCO2 is above the upper alarm limit and may indicate a real patient ventilation issue, airway issue, rebreathing condition, incorrect alarm limit, sampling accessory problem, or unreliable EtCO2 measurement.
Because elevated EtCO2 can reflect patient compromise, this alarm should always be treated as a patient safety alarm first. Equipment troubleshooting should only begin after the patient has been assessed and safe monitoring is maintained.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Treat the EtCO2 HIGH ALARM as a patient safety alarm first, not an equipment problem.
- Confirm the patient is being assessed immediately by clinical staff.
- Do not silence, ignore, or troubleshoot the monitor as the first action if the patient is connected.
- Make sure appropriate clinical staff evaluate the patient for respiratory, ventilation, airway, or clinical condition changes.
Verify the Reported Alarm and Behavior
- Confirm the monitor is displaying EtCO2 HIGH ALARM.
- Observe the displayed EtCO2 value, capnogram waveform, upper alarm limit, and alarm timing.
- Check whether the alarm occurs continuously, intermittently, during patient movement, after a circuit change, during transport, or after alarm limits were adjusted.
- Ask staff whether the alarm appeared suddenly or followed a known setup, accessory, or patient condition change.
Confirm Clinical Response Before Equipment Troubleshooting
- Make sure respiratory therapy, nursing, anesthesia, or the responsible clinical team has evaluated the patient and ventilation setup.
- Do not remove the monitor from use or disconnect monitoring until the patient is safely assessed.
- If monitoring must be interrupted, make sure alternate monitoring is available first.
Check the Alarm Limit Setting
- Review the upper EtCO2 alarm limit shown on the monitor.
- Confirm the limit is appropriate for the patient population, care area, and clinical situation.
- Look for accidental alarm limit changes, incorrect patient profile, or a limit set too low.
- Only authorized clinical staff should change patient alarm limits according to facility policy.
Inspect the CO2 Sampling Line or Airway Adapter
- Check the sampling line, FilterLine, airway adapter, or nasal/oral cannula connection.
- Look for kinks, partial obstruction, moisture, contamination, loose connections, incorrect placement, or damaged disposable accessories.
- Replace the disposable CO2 sampling accessory if it is contaminated, wet, restricted, damaged, expired, or questionable.
- Confirm the sampling accessory is appropriate for the patient and monitoring setup.
Check the Capnogram Waveform
- Observe whether the capnogram waveform appears clinically reasonable, square, elevated, distorted, erratic, or inconsistent with the patient presentation.
- Compare the waveform behavior with the EtCO2 number and alarm condition.
- A stable waveform that matches the elevated EtCO2 value may support a true measured condition.
- An unstable, distorted, or inconsistent waveform may suggest accessory, setup, or sampling artifact.
Check for Rebreathing or External Setup Issues
- From a Clinical Engineering support standpoint, inspect only the external setup that is safe to verify.
- Check tubing routing, sampling line placement, airway adapter position, exhaust or scavenging setup if applicable, and obvious connection errors.
- Do not adjust ventilator, anesthesia, or respiratory settings unless that is within your role and facility policy.
- Escalate any suspected ventilation or rebreathing issue to the responsible clinical team.
Compare With Another Monitor or Measurement Source if Needed
- If the patient is stable and clinical staff suspect a monitor issue, compare the EtCO2 reading using another known-good capnography monitor, module, or approved measurement source.
- Use a known-good sampling line or accessory during comparison.
- If both devices show similar values, the issue may be patient-related, setup-related, or alarm-limit-related.
- If the Capnostream 20p reading does not agree with a known-good comparison, remove it from service for evaluation.
Check Power, Connections, and Physical Condition
- Confirm the monitor has stable power and is not switching unexpectedly between AC and battery power.
- Check that external connections are secure.
- Inspect the monitor for liquid intrusion, cracked ports, damaged connectors, impact damage, contamination, or signs of overheating.
- Do not continue using a monitor with obvious physical damage or unreliable EtCO2 behavior.
Remove From Service if Equipment Fault Is Suspected
- If the alarm continues with a stable patient, correct alarm limits, known-good sampling accessories, and comparison testing suggests the monitor is reading incorrectly, remove the monitor from service.
- Label the device clearly as Out of Service.
- Send it for repair or bench evaluation according to facility procedure.
If the Problem Persists
If clinical causes, alarm settings, sampling accessories, power, external connections, and setup issues have been ruled out, the persistent alarm may indicate a device measurement issue or internal CO2 monitoring problem.
Remove the Capnostream 20p from service, label it Out of Service, and send it for repair or bench evaluation.
Knowing when to stop is proper troubleshooting. Do not keep troubleshooting on an active patient or continue using a monitor with unresolved EtCO2 alarm concerns.
Clinical Use Tip
Never troubleshoot this alarm on an active patient as if it is only an equipment issue. A high EtCO2 alarm can indicate a real ventilation problem. Move the patient to a backup monitor or approved alternate monitoring method first if the Capnostream 20p must be removed from patient use.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Capnostream 20p reported with EtCO2 HIGH ALARM. User stated EtCO2 value was above the upper alarm limit during patient monitoring."
Cause
What was observed during troubleshooting.
Example:
"Alarm confirmed. Clinical evaluation required first due to elevated EtCO2 reading. External checks included alarm limit review, sampling line inspection, FilterLine replacement, waveform observation, power check, and physical inspection."
Resolution
What action was taken.
Example:
"Verified alarm behavior and confirmed clinical staff response. Replaced questionable sampling accessory and checked external connections. If alarm cleared with correct setup and clinical staff approved, device returned to service. If alarm persisted with known-good accessories or questionable EtCO2 readings, device removed from service for bench evaluation."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
An EtCO2 HIGH ALARM must be treated as a patient condition first. Clinical Engineering should support safe troubleshooting by confirming the alarm, checking simple external causes, verifying accessories and alarm settings, and escalating the device if the reading appears unreliable. Clear documentation protects the patient, the staff, and the equipment history.
That is successful troubleshooting.