Medtronic Capnostream 20p

Alarm: EtCO2 LOW ALARM

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot the EtCO2 LOW ALARM on a Medtronic Capnostream 20p capnography monitor. This alarm indicates that the measured end-tidal CO2 value is below the lower alarm limit. It may reflect a real patient condition, poor ventilation, airway or circuit problems, low perfusion, apnea risk, or a false low reading caused by sampling line, cannula, airway adapter, or connection issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Whether the Reading Matches the Patient Condition

Check the Lower EtCO2 Alarm Limit

Inspect the CO2 Sampling Line or Airway Adapter

Confirm Proper Patient Interface Placement

Check for Leaks or Disconnections in the Breathing Circuit

Evaluate the CO2 Waveform

Swap Known-Good Accessories

Check the Monitor Connection

Test the Monitor Away From the Patient

If the Problem Persists

If the EtCO2 LOW ALARM continues after the patient has been assessed, alarm limits are verified, the sampling line and accessories are replaced, connections are checked, and the issue remains during bench evaluation, external causes have been ruled out.

Persistent false low EtCO2 readings may indicate a device sampling, sensor, pneumatic, calibration, or internal measurement problem.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot a low EtCO2 alarm as an equipment issue while the patient is actively relying on the monitor. Move the patient to a backup device or alternate monitoring method first, then troubleshoot the Capnostream 20p only after patient monitoring remains safe.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p displayed EtCO2 LOW ALARM during patient monitoring. Staff reported EtCO2 value below the lower alarm limit with concern for low or inaccurate CO2 reading."

Cause

What was observed during troubleshooting.

Example:
"Patient was assessed by clinical staff. Lower EtCO2 alarm limit, CO2 sampling line, patient interface placement, airway/circuit connections, and monitor inlet connection were checked. Issue was caused by a displaced/kinked CO2 sample line, which prevented accurate exhaled CO2 sampling."

Resolution

What action was taken.

Example:
"Replaced/repositioned CO2 sampling line and confirmed stable EtCO2 numeric value, visible waveform, and no recurring EtCO2 LOW ALARM. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

The EtCO2 LOW ALARM must be treated as a patient safety alarm first and an equipment issue second. Clinical Engineering should verify the alarm, rule out simple external causes, confirm accessory and sampling integrity, and escalate when the monitor cannot be trusted. Clear CCR documentation protects the patient, the technician, and the device history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide