Medtronic Capnostream 20p

Alarm: FiCO2 RATE HIGH ALARM

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot a FiCO2 RATE HIGH ALARM on a Medtronic Capnostream 20p capnography monitor. This alarm indicates the inspired CO2 rate is above the upper alarm limit and may be related to patient condition, ventilation setup, CO2 rebreathing, sampling line placement, accessory issues, or monitor measurement problems.

This alarm should be treated as a patient safety concern first. Clinical Engineering should not assume the device is faulty until the patient, breathing circuit, sampling setup, alarm settings, and external accessories have been checked.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm the Patient Is Not Actively Rebreathing CO2

Check the Sampling Line or FilterLine Connection

Inspect the Patient Interface or Airway Adapter

Check for Environmental or Setup-Related Causes

Review Alarm Limits and Configuration

Compare With Other Patient Parameters

Test With Known-Good Accessories

Remove the Monitor From Service if the Alarm Appears False and Persists

If the Problem Persists

If patient condition, ventilation setup, sampling line placement, accessories, alarm limits, and environmental causes have been ruled out, the alarm may indicate a device measurement problem or internal CO2 monitoring issue.

Remove the device from service, label it Out of Service, and send it for repair or bench evaluation.

Do not continue using a monitor that repeatedly reports a high FiCO2 condition without a confirmed clinical or setup-related cause.

Clinical Use Tip

Never troubleshoot this alarm on an active patient as if it is only an equipment issue. Move the patient to a backup device first if monitoring is required and the alarm cannot be quickly explained. A FiCO2 high alarm may reflect a real ventilation or rebreathing concern.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p displayed FiCO2 RATE HIGH ALARM during patient monitoring. Staff reported inspired CO2 reading above the upper alarm limit."

Cause

What was observed during troubleshooting.

Example:
"Patient safety confirmed with clinical staff. Sampling line, airway interface, alarm limit, waveform quality, and accessory condition checked. Alarm appeared related to sampling setup/accessory condition."

Resolution

What action was taken.

Example:
"Replaced CO2 sampling accessory and verified stable readings and waveform. Alarm cleared during testing. Monitor returned to service. If alarm had persisted, unit would have been removed from service for bench evaluation."

Helpful Details to Include (If Known)

Final Thought

A FiCO2 RATE HIGH ALARM should be handled with patient safety first. Rule out patient condition, ventilation setup, sampling accessories, alarm configuration, and environmental causes before suspecting monitor failure. If the alarm persists without a clear external or clinical cause, remove the device from service and document the work clearly using the CCR method.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide