Medtronic Capnostream 20p

Alarm: FILTERLINE BLOCKAGE

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot the FILTERLINE BLOCKAGE alarm on a Medtronic Capnostream 20p capnography monitor. This alarm indicates that the FilterLine is kinked, clogged, or restricted, which can prevent accurate CO2 sampling and interrupt EtCO2 monitoring.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Check the FilterLine Routing

Disconnect and Reconnect the FilterLine

Inspect the Airway Adapter or Patient Connection

Replace the FilterLine

Check for Use-Related or Environmental Causes

Test With a Known-Good FilterLine

If the Problem Persists

If the FILTERLINE BLOCKAGE alarm continues after the FilterLine has been reconnected, the airway adapter or patient connection has been checked, routing has been corrected, and a new compatible FilterLine has been installed, common external causes have been ruled out.

A persistent alarm may indicate a device-side sampling problem, internal restriction, pump or sampling pathway issue, or monitor fault.

Remove the Capnostream 20p from service, label it Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot this alarm on an active patient without ensuring the patient is safely monitored first. Move the patient to a backup monitor or alternate monitoring method before troubleshooting the Capnostream 20p.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p displayed FILTERLINE BLOCKAGE alarm during use and EtCO2 monitoring was interrupted."

Cause

What was observed during troubleshooting.

Example:
"FilterLine was kinked at the patient connection, restricting CO2 sampling."

Resolution

What action was taken.

Example:
"Inspected FilterLine routing, reconnected the FilterLine, checked the airway adapter, replaced the FilterLine, and verified normal CO2 sampling with no recurring FILTERLINE BLOCKAGE alarm."

Helpful Details to Include (If Known)

Final Thought

A FILTERLINE BLOCKAGE alarm should be treated as a sampling-path problem first. Start with patient safety, confirm the alarm, inspect the FilterLine and airway adapter, replace the disposable line when needed, and escalate only after external causes have been ruled out. Clear documentation helps show whether the problem was accessory-related, use-related, or likely device-related.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide