Medtronic Capnostream 20p

Alarm: IPI LOW ALARM

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot an IPI LOW ALARM on a Medtronic Capnostream 20p capnography monitor. The alarm indicates that the Integrated Pulmonary Index is below the lower alarm limit and may represent a change in patient status. This should be treated as a patient safety alarm first, with equipment and accessory checks performed only after the patient has been evaluated.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Review Other Monitored Parameters

Confirm Alarm Limits and Patient Setup

Check the CO2 Sampling Setup

Check the SpO2 Sensor and Patient Connection

Check Cables, Accessories, and Connections

Assess for Motion or Environmental Causes

If the Problem Persists

If the IPI LOW ALARM continues after the patient has been evaluated, alarm settings are confirmed, and known-good CO2 and SpO2 accessories have been used, external causes have been ruled out.

Remove the monitor from service. Label it Out of Service and send it for repair or bench evaluation. Persistent IPI alarms with unreliable parameter readings may indicate a device, module, connector, or signal-processing issue.

Clinical Use Tip

Never troubleshoot this alarm on an active patient until clinical staff have assessed the patient and confirmed it is safe to continue. If the monitor or accessories cannot provide reliable readings, move the patient to a backup monitor first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p reported IPI LOW ALARM during patient monitoring. Staff reported low IPI indication and requested evaluation of monitor and accessories."

Cause

What was observed during troubleshooting.

Example:
"Patient status alarm required clinical evaluation. CO2 sampling line and SpO2 sensor were checked for placement, connection, damage, and signal quality. Alarm condition was supported by monitored parameters / caused by unstable accessory signal / could not be duplicated after accessory correction."

Resolution

What action was taken.

Example:
"Patient care team evaluated patient. Verified alarm behavior, reviewed EtCO2, respiration rate, SpO2, pulse rate, and IPI trend. Checked sampling line, SpO2 sensor, cables, and alarm settings. Corrected accessory issue and returned monitor to service, or removed monitor from service for bench evaluation if alarm persisted."

Helpful Details to Include (If Known)

Final Thought

An IPI LOW ALARM should be handled as a patient safety event first and an equipment issue second. Clinical Engineering should confirm the alarm, check simple external causes, verify accessory performance, and escalate only after patient condition and common causes have been addressed.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide