Medtronic Capnostream 20p

Alarm: No breath xxx seconds

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot the No breath xxx seconds alarm on a Medtronic Capnostream 20p capnography monitor. This alarm indicates that no valid breath has been detected for the displayed number of seconds.

This condition may be caused by a real patient breathing issue, airway problem, disconnected sampling line, poor patient interface placement, blocked CO₂ accessory, incorrect setup, or a device-side CO₂ sampling problem. This alarm must be treated as a patient safety alarm first.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Clinical Response Before Equipment Checks

Check the CO₂ Sampling Line and Airway Connection

Check the Patient Interface Placement

Verify the Capnogram Waveform and EtCO₂ Reading

Replace the Sampling Line or CO₂ Accessory

Check the Monitor-Side Connection

Check Alarm Settings and Patient Configuration

Isolate Accessory or Patient Setup from Device Failure

Stop When Resolved

If the Problem Persists

If the patient has been assessed, the sampling line and patient interface are correct, compatible accessories have been replaced, the monitor connection is secure, and the Capnostream 20p still displays No breath xxx seconds, external and common causes have been ruled out.

Persistent no-breath detection despite a valid setup may indicate an internal sampling, pump, sensor, connector, or CO₂ measurement problem.

Remove the monitor from service, label it Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot this alarm on an active patient as if it is only an equipment issue. A No breath alarm can represent apnea, airway obstruction, respiratory depression, or ventilation failure.

Move the patient to a backup monitor or alternate monitoring method first if the device cannot be trusted or if troubleshooting could interrupt patient monitoring.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p displayed No breath xxx seconds alarm and staff reported no valid breath was being detected by the monitor."

Cause

What was observed during troubleshooting.

Example:
"Patient safety alarm verified. Clinical staff assessed patient. CO₂ sampling line and patient interface were checked for disconnection, blockage, moisture, poor placement, or accessory failure."

Resolution

What action was taken.

Example:
"Replaced sampling line and confirmed stable CO₂ waveform, EtCO₂ reading, and cleared alarm. Monitor returned to service after normal operation was verified."

Helpful Details to Include (If Known)

Final Thought

The No breath xxx seconds alarm must be treated as a patient safety alarm before it is treated as a device problem. Confirm the patient, verify the sampling path, check the waveform, replace simple accessories, and escalate the monitor only after external causes are ruled out.

Good troubleshooting protects the patient, avoids unnecessary repair work, and creates a clear service history for the device.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide