Medtronic Capnostream 20p

Alarm: PULSE NOT FOUND

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot the PULSE NOT FOUND alarm on the Medtronic Capnostream 20p capnography monitor. This alarm indicates that the monitor cannot detect a pulse from the SpO2/pulse monitoring function. Because this may reflect a real patient condition, poor sensor placement, poor perfusion, motion artifact, a bad sensor, a cable issue, or a monitor-side fault, patient safety must be confirmed before treating this as an equipment problem.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Reposition the SpO2/Pulse Sensor

Check the Patient Site and Signal Quality

Inspect the Sensor and Cable Connection

Swap the Sensor or Accessory

Compare With Another Monitor if Needed

Confirm Stable Operation Before Return to Use

If the Problem Persists

If patient condition has been assessed, sensor placement has been corrected, the sensor site has been checked, known-good compatible accessories have been tested, and the PULSE NOT FOUND alarm remains, remove the Capnostream 20p from service.

Label the device Out of Service and send it for repair or bench evaluation. Persistent pulse detection failure after external causes are ruled out may indicate a SpO2 input, accessory interface, or internal monitoring fault.

Clinical Use Tip

Never troubleshoot this alarm on an active patient while reliable monitoring is still needed. Move the patient to a backup monitor or approved pulse oximeter first, then troubleshoot the Capnostream 20p only after the patient is safe and monitoring has been restored.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p displayed PULSE NOT FOUND alarm and staff reported no detectable pulse reading from the SpO2/pulse sensor."

Cause

What was observed during troubleshooting.

Example:
"Patient safety confirmed by clinical staff. Sensor placement, patient site, cable connection, and accessory condition were checked. Alarm cleared after SpO2 sensor was repositioned/replaced."

Resolution

What action was taken.

Example:
"Repositioned SpO2 sensor, inspected cable connection, and tested with known-good compatible accessory. Verified stable pulse detection and returned monitor to service."

Helpful Details to Include (If Known)

Final Thought

The PULSE NOT FOUND alarm must be handled as a patient safety concern first and an equipment issue second. Confirm the patient is safe, restore reliable monitoring, then work through sensor placement, patient site, cable connection, accessory condition, and comparison testing before escalating the monitor for repair. Clear CCR documentation shows what was checked, what corrected the problem, and whether the device was safe to return to service.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide