Medtronic Capnostream 20p

Alarm: PULSE RATE HIGH ALARM

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot a PULSE RATE HIGH ALARM on the Medtronic Capnostream 20p when the displayed pulse rate is above the upper alarm limit. This may represent a real patient condition requiring immediate clinical attention, or it may be caused by alarm limit settings, poor SpO2 sensor placement, motion artifact, a damaged sensor, a faulty cable, poor signal quality, or a monitor-related issue.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm the Pulse Rate Against Another Source

Check Alarm Limit Settings

Inspect the SpO2 Sensor Placement

Check for Motion Artifact

Inspect the SpO2 Cable and Connections

Swap the SpO2 Sensor or Cable

Check Signal Quality and Waveform

Perform a Basic Functional Check

If the Problem Persists

If the PULSE RATE HIGH ALARM continues after the patient condition has been ruled out, alarm limits are appropriate, sensor placement has been checked, motion artifact has been reduced, and known-good SpO2 accessories have been tested, external causes have been reasonably ruled out.

Remove the Capnostream 20p from service. Label it Out of Service. Send the device for repair or bench evaluation.

Persistent false pulse rate alarms should not be ignored because they can cause alarm fatigue, distract staff, or interfere with recognition of true patient events.

Clinical Use Tip

Never troubleshoot this alarm on an active patient until clinical staff confirms the patient is safe. If monitoring must continue, move the patient to a backup device first before removing accessories, changing cables, or testing the monitor.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p displayed PULSE RATE HIGH ALARM during patient monitoring. Staff reported pulse rate reading above the upper alarm limit."

Cause

What was observed during troubleshooting.

Example:
"Patient assessment was required first. Equipment checks found unstable SpO2 signal related to poor sensor contact and movement artifact. Alarm limit and cable connection were also checked."

Resolution

What action was taken.

Example:
"Confirmed reported alarm, verified alarm settings, inspected SpO2 sensor placement, checked cable connection, and tested with known-good SpO2 accessory. Readings stabilized after sensor replacement and monitor passed functional check."

Helpful Details to Include (If Known)

Final Thought

A PULSE RATE HIGH ALARM should always be treated as a patient safety alarm first. Clinical Engineering should verify the alarm, support clinical staff, rule out external causes, check accessories and settings, and escalate the monitor only after common causes have been eliminated. Clear CCR documentation protects the patient, the technician, and the facility.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide