Medtronic Capnostream 20p

Alarm: RR HIGH ALARM

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot RR HIGH ALARM on a Medtronic Capnostream 20p capnography monitor. This alarm indicates the respiration rate is above the upper alarm limit. It may be caused by an actual elevated respiratory rate, patient condition, alarm limit settings, sampling issues, waveform artifact, movement, or false alarm behavior.

Because a high respiration rate can reflect respiratory distress or another clinical change, this alarm should be treated as a patient safety issue first and an equipment issue second.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Compare the Displayed Respiratory Rate to the Patient

Confirm Alarm Limits Are Appropriate

Check the CO2 Sampling Line or Airway Connection

Inspect the Capnogram Waveform

Check for Patient Movement or Artifact

Replace the Disposable Sampling Accessory if False Readings Are Suspected

Check the Monitor Setup and Patient Type

Test the Monitor Away From the Patient if a Device Issue Is Suspected

Stop When Resolved

If the Problem Persists

If the RR HIGH ALARM continues after patient condition, alarm limits, sampling line placement, waveform quality, movement artifact, and known-good accessories have been checked, remove the monitor from service.

Label the unit Out of Service and send it for repair or bench evaluation. Persistent false respiratory rate alarms after external causes are ruled out may indicate a sampling, sensor, processing, or device-related failure.

Clinical Use Tip

Never troubleshoot a capnography alarm on an active patient without clinical staff involved. A high respiratory rate alarm should be treated as a patient condition first, not a device problem. Move the patient to a backup monitor first if the alarm cannot be trusted or if monitoring must continue during troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p reported RR HIGH ALARM during patient monitoring. Staff stated respiration rate was above the upper alarm limit."

Cause

What was observed during troubleshooting.

Example:
"Verified RR HIGH ALARM. Patient assessment required first. Checked upper RR alarm limit, sampling line position, waveform quality, accessory condition, and movement artifact. Alarm appeared related to patient respiratory rate or alarm limit condition rather than confirmed device failure."

Resolution

What action was taken.

Example:
"Clinical staff notified for patient assessment. Sampling line and connections checked. Alarm limit reviewed by clinical staff. Monitor function verified with stable capnogram waveform. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

The RR HIGH ALARM should always be treated as a possible patient condition first. Clinical Engineering should verify the alarm, check simple external causes, confirm waveform quality, and only escalate the device after patient condition, alarm settings, and accessories have been ruled out. Clear CCR documentation protects the patient, the technician, and the equipment history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide