Medtronic Capnostream 20p

Alarm: RR LOW ALARM

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

This guide helps troubleshoot the RR LOW ALARM on a Medtronic Capnostream 20p capnography monitor. This alarm indicates that the measured respiration rate is below the lower alarm limit. It may reflect a true patient condition, poor breath detection, an accessory issue, incorrect alarm settings, or a monitoring setup problem.

Because a low respiratory rate can indicate respiratory depression, apnea risk, oversedation, airway obstruction, or clinical deterioration, this alarm must be treated as a patient safety alarm before it is treated as an equipment problem.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Clinical Response Before Equipment Troubleshooting

Check the Alarm Limit Setting

Inspect the Sampling Line or Cannula

Check Patient Interface Placement

Check the CO2 Waveform

Replace the Sampling Accessory

Check for Environmental or Use-Related Causes

Compare With Another Monitor or Known-Good Setup

Remove From Service if the Alarm Appears False and Cannot Be Corrected

If the Problem Persists

If the RR LOW ALARM continues after patient condition, alarm settings, sampling line, accessory placement, waveform quality, environmental factors, and known-good accessories have been checked, external and common causes have been ruled out.

Persistent false low respiratory rate alarms may indicate a CO2 sampling, detection, sensor, or internal monitor issue.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot this alarm on an active patient while ignoring the patient condition. A low respiration rate can be a true clinical emergency. Move the patient to a backup monitor first if monitoring must continue while the Capnostream 20p is evaluated.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Capnostream 20p displayed RR LOW ALARM during patient monitoring. Staff reported respiration rate was below the lower alarm limit or alarm appeared despite observed breathing."

Cause

What was observed during troubleshooting.

Example:
"Verified RR LOW ALARM. Clinical staff assessed patient condition. Checked lower RR alarm limit, sampling line, cannula placement, CO2 waveform, and accessory condition. Alarm was either consistent with patient respiratory status or suspected false after external causes were ruled out."

Resolution

What action was taken.

Example:
"Corrected sampling line/cannula placement, replaced CO2 sampling accessory, verified stable waveform and respiration rate reading, and returned monitor to service. If alarm persisted with known-good accessories, removed device from service and sent for bench evaluation."

Helpful Details to Include (If Known)

Final Thought

The RR LOW ALARM must be treated as a possible patient emergency before assuming equipment failure. Clinical Engineering should confirm the alarm, verify patient safety response, check simple external causes, inspect sampling accessories, review waveform quality, and escalate only when the monitor remains unreliable after common causes are ruled out.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide