On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting CO2 ERROR or loss of CO₂ monitoring before removing the Capnostream 20p from service.
Step-by-Step Troubleshooting
Ensure Patient Safety First
If the Capnostream 20p is actively monitoring a patient and CO₂ monitoring is needed, move the patient to another working monitor or use an approved alternate monitoring method before troubleshooting.
Expected outcome: The patient remains safely monitored.
Why it matters: A CO₂ function failure may prevent reliable EtCO₂, respiration rate, waveform, and apnea monitoring.
Verify the Reported Error
Confirm that the displayed message is CO2 ERROR or that the CO₂ function is unavailable. Note whether the issue appears during startup, after connecting a sampling line, during monitoring, or intermittently.
Expected outcome: The exact failure condition is confirmed.
Why it matters: This separates a true CO₂ function failure from a simple sampling line, setup, or accessory problem.
Check Whether CO₂ Values and Waveform Are Missing
Look for a missing capnogram waveform, dashed EtCO₂ value, missing respiration rate, or disabled CO₂ alarms.
Expected outcome: The scope of the CO₂ failure is understood.
Why it matters: A complete loss of CO₂ data points more strongly toward a device or module issue.
Confirm the Monitor Is Powered Correctly
Connect the Capnostream 20p to a known-good AC outlet. Verify the power cord is fully seated and that the monitor powers on normally without low battery or power-related alarms.
Expected outcome: The monitor runs from stable AC power.
Why it matters: Low power, battery issues, or unstable AC connection can cause abnormal startup or functional errors.
Power Cycle the Monitor
Turn the monitor off, wait briefly, then restart it while connected to AC power. Observe whether the CO2 ERROR returns immediately during startup.
Expected outcome: The monitor starts normally and the CO₂ function becomes available.
If resolved: Document the reset and return the device to service only after functional checks pass.
Why it matters: Some errors may clear after a controlled restart, but recurring errors need further evaluation.
Inspect the CO₂ Sampling Line Connection
Remove and reconnect the sampling line. Make sure it is fully seated, not cross-threaded, loose, damaged, kinked, wet, contaminated, or blocked.
Expected outcome: The line connects securely and allows proper sampling.
Why it matters: A bad or obstructed sampling line can mimic CO₂ monitoring failure or prevent valid breath detection.
Replace the Sampling Line With a Known-Good Compatible Line
Use a new or verified compatible Microstream sampling line appropriate for the patient setup.
Expected outcome: CO₂ sampling resumes, waveform appears, and EtCO₂/respiration rate display normally.
If resolved: The original sampling line was likely defective, blocked, expired, contaminated, or incompatible. Stop troubleshooting.
Check for Moisture, Contamination, or Visible Damage at the CO₂ Inlet
Inspect the sampling port area externally. Do not insert tools into the port unless following approved site procedure. Look for liquid exposure, residue, cracks, loose fit, or physical damage.
Expected outcome: The inlet appears clean, dry, and physically intact.
Why it matters: Moisture or contamination can affect the sampling system and may require bench service.
Allow Warm-Up or Initialization to Complete
After startup, give the monitor time to complete its normal initialization. Watch whether the error clears or remains active.
Expected outcome: CO₂ function becomes available after normal startup.
Why it matters: Calling a failure too early can lead to unnecessary repair evaluation.
Test With a Simple Functional Check
If your site procedure allows it, connect a known-good sampling line and verify that the monitor detects CO₂ using an approved test source or simulator. Do not use patient breath testing unless permitted by your facility policy.
Expected outcome: A stable waveform and EtCO₂ value appear.
If resolved: The monitor passes basic CO₂ function testing and may be returned to service after documentation.
Why it matters: A functional check confirms whether the CO₂ system is actually operating.
Check Alarm and Parameter Behavior After CO₂ Returns
Confirm that EtCO₂, respiration rate, waveform, and apnea/no-breath alarm behavior are available and not disabled by setup or configuration.
Expected outcome: CO₂ monitoring and related alarms operate as expected.
Why it matters: Restoring a waveform is not enough if the related monitoring and alarms are still unavailable.
Repeat the Failure Check After Movement
Gently move the monitor, power cord, and sampling line connection while observing the screen. Do not stress the connectors.
Expected outcome: The CO2 ERROR does not return.
Why it matters: Intermittent connector or internal sampling system issues may only appear during movement or transport.
Do Not Continue Using the Monitor if CO₂ Failure Repeats
If CO2 ERROR returns after power cycling, accessory replacement, and basic external checks, remove the unit from service.
Expected outcome: The device is not used for patient monitoring until evaluated.
Why it matters: A recurring CO₂ function failure can compromise respiratory monitoring and alarm reliability.
If the Problem Persists
If the CO2 ERROR remains after checking power, restarting the monitor, replacing the sampling line, inspecting the inlet, and performing a basic functional check, common external causes have been ruled out.
The issue is likely internal to the CO₂ measurement system, sampling pump path, sensor/module electronics, internal tubing, or related circuitry.
The device must be removed from service, labeled Out of Service, and sent for repair, vendor service, or bench evaluation according to your facility process.
Knowing when to stop is proper troubleshooting. Do not keep returning a capnography monitor to service if the CO₂ function cannot be verified.
Clinical Use Tip
Do not troubleshoot a CO₂ function failure on an actively monitored patient unless another approved monitoring method is already in place. Move the patient to a backup monitor first if CO₂ monitoring is needed for therapy continuity. A Capnostream 20p with CO₂ failure may still power on, but it should not be trusted for EtCO₂, respiration rate, apnea, or no-breath monitoring until the CO₂ function is verified.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported CO2 ERROR on the Medtronic Capnostream 20p and stated that EtCO₂ and respiration rate were not displaying during monitoring."
Cause
What was observed during troubleshooting.
Example:
"Verified CO2 ERROR during startup and confirmed no CO₂ waveform using a known-good sampling line and stable AC power."
Resolution
What action was taken.
Example:
"Removed monitor from service, labeled Out of Service, and sent unit for bench evaluation due to unresolved CO₂ function failure."
Helpful Details to Include (If Known)
- Whether the unit was on AC power or battery
- Outlet tested or alternate outlet used
- Power cord condition
- Exact displayed error message
- Whether error occurred at startup or during monitoring
- Sampling line replaced with known-good compatible line
- CO₂ inlet inspected for damage, moisture, or contamination
- CO₂ waveform present or missing
- EtCO₂ value present, dashed, unstable, or absent
- Respiration rate present or missing
- Alarm behavior before and after troubleshooting
- Accessories swapped during troubleshooting
- Power behavior during startup, battery use, and AC connection
- Environmental factors such as liquid exposure, transport damage, contamination, or recent cleaning
- Indicator lights, alarms, or display behavior observed during startup
- Any unusual sound from sampling pump area
- Any unusual heat, smell, liquid exposure, or physical damage
- Final device status: returned to service, removed from service, or sent for repair
Final Thought
A CO2 ERROR on a Capnostream 20p should be treated as a patient monitoring reliability issue, not just a nuisance message. Clinical Engineering should rule out power, setup, sampling line, and external connection problems first, then escalate when CO₂ function cannot be verified. Clear CCR documentation protects the patient, the technician, and the equipment record.
That is successful troubleshooting.