Medtronic Capnostream 20p

No Breath Alarm or Apnea Alarm During Monitoring

On this page

Asset Type

Capnography Monitor

Manufacturer

Medtronic

Model

Capnostream 20p

What This Guide Helps With

Troubleshooting no breath or apnea alarms caused by patient condition, sampling line issues, setup problems, or unreliable breath detection.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Treat a No Breath or apnea alarm as a patient safety alarm first, not an equipment problem. Confirm clinical staff are assessing the patient immediately.

Expected outcome: The patient is checked for apnea, airway obstruction, respiratory depression, disconnection, or ventilation failure.

Why it matters: A no breath alarm may indicate the patient is not breathing or that ventilation is not being detected.

Do Not Troubleshoot on an Unstable Active Patient

If the patient requires continued monitoring, move monitoring to another approved capnography monitor or alternate monitoring method before troubleshooting the Capnostream 20p.

Expected outcome: Patient monitoring continues safely while the suspect device is evaluated.

If this resolves the immediate safety concern, continue troubleshooting only when safe.

Verify the Reported Alarm

Confirm the monitor is displaying a No Breath, apnea, or related breath detection alarm. Note the alarm wording, alarm delay time, waveform behavior, EtCO2 value, and respiration rate display.

Expected outcome: The exact alarm and monitor behavior are confirmed.

Why it matters: This separates a true apnea/no breath alarm from a staff report, nuisance alarm, or unrelated CO2 alarm.

Check for a Valid CO2 Waveform

Look for a normal capnogram waveform. Check whether the waveform is absent, flat, very small, intermittent, noisy, or delayed.

Expected outcome: A clear waveform should appear with each detected breath.

If the waveform returns and breath detection normalizes, document the correction and stop.

Inspect the Sampling Line or Cannula Connection

Confirm the Microstream sampling line is fully connected to the monitor and properly connected to the patient or airway circuit.

Expected outcome: The sampling line is secure at both ends.

Why it matters: A loose or disconnected line can prevent the monitor from detecting exhaled CO2.

Check for Kinks, Occlusion, Moisture, or Contamination

Inspect the full length of the sampling line. Look for bends, pinching under bed rails, fluid, secretions, crushed tubing, or a blocked filter area.

Expected outcome: The sampling path is open and dry.

If a kink or blockage is corrected and the alarm clears, document the finding and stop.

Replace the Sampling Line

Install a new compatible sampling line or cannula. Do not reuse a visibly contaminated, wet, damaged, or suspect line.

Expected outcome: CO2 waveform and respiration rate return if the original line was defective or blocked.

Why it matters: Sampling lines are common external failure points and should be ruled out before suspecting the monitor.

Confirm the Correct Patient Setup

Verify the monitor setup matches the clinical use case, such as adult, pediatric, or neonatal monitoring if applicable. Confirm the correct sampling accessory is being used for nasal/oral, intubated, or non-intubated monitoring.

Expected outcome: The monitor and accessory setup match the patient and monitoring method.

Why it matters: Incorrect setup can cause poor breath detection or nuisance apnea alarms.

Check Patient Placement and Cannula Position

Confirm the cannula is positioned correctly at the nares and mouth area if monitoring a non-intubated patient. Check whether the patient is mouth breathing, talking, moving, receiving oxygen, or has the cannula displaced.

Expected outcome: The sampling line captures exhaled breath consistently.

If repositioning restores waveform and RR detection, document the correction and stop.

Check Ventilator or Airway Circuit Connection if Intubated

For intubated or ventilated patients, confirm the sampling adapter is connected correctly in the breathing circuit and is not loose, cracked, blocked, or filled with condensation.

Expected outcome: Exhaled CO2 reaches the sampling line with each breath.

Why it matters: Circuit connection issues can mimic apnea even when the patient or ventilator is breathing.

Review Alarm Settings and Apnea Delay

Confirm the apnea/no breath alarm delay and respiratory alarm settings are appropriate for the patient population and clinical area. Do not change alarm settings without clinical approval.

Expected outcome: Alarm limits and delay settings match the clinical monitoring need.

Why it matters: An overly short apnea delay may create nuisance alarms, while an overly long delay can delay response.

Check for Low EtCO2 or Shallow Breathing

Observe whether the patient has very low EtCO2, shallow respirations, weak waveform amplitude, or irregular breathing. Confirm clinical staff are aware.

Expected outcome: The alarm is understood as either a clinical condition or a detection issue.

Why it matters: The monitor may alarm because it cannot detect valid breaths, even if slight airflow is present.

Power Cycle Only When Safe

If the patient is no longer connected or monitoring has been transferred, power cycle the Capnostream 20p and allow it to restart normally.

Expected outcome: The monitor boots without CO2 function errors and resumes normal sampling behavior.

Do not power cycle during active patient monitoring unless clinical staff confirm it is safe.

Test With a Known-Good Setup

Use a known-good sampling line and a safe functional check method approved by your department. Confirm the monitor detects CO2 waveform and respiration rate consistently.

Expected outcome: The Capnostream 20p detects breaths reliably with known-good accessories.

If the monitor works with known-good accessories, the original line, setup, or patient connection was likely the cause.

Check for Related Device Errors

Look for CO2 ERROR, pump-related messages, blocked line messages, startup errors, unusual noise from the internal pump, or failure to draw sample flow.

Expected outcome: No internal CO2 system fault is present.

Why it matters: Persistent no breath alarms with good external setup may indicate an internal sampling or CO2 module issue.

If the Problem Persists

If patient condition, sampling line, accessory connection, cannula placement, alarm setup, and known-good accessory testing have been ruled out, the issue is likely internal to the monitor or CO2 sampling system.

Remove the Capnostream 20p from service, label it Out of Service, and send it for bench evaluation or repair. Do not return the monitor to patient use if it cannot reliably detect breaths, display a stable waveform, or maintain CO2 monitoring.

Knowing when to stop is proper troubleshooting. A no breath or apnea alarm can represent a serious patient condition, so the device should not be trusted until the cause is confirmed.

Clinical Use Tip

Do not troubleshoot a no breath or apnea alarm as a simple equipment complaint while the patient is still connected. Confirm the patient is breathing and being clinically assessed first. If monitoring is still needed, move the patient to another working monitor before continuing equipment checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported the Capnostream 20p was alarming No Breath during patient monitoring."

Cause

What was observed during troubleshooting.

Example:
"Found the CO2 sampling line kinked under the bed rail, causing intermittent waveform loss and unreliable respiration detection."

Resolution

What action was taken.

Example:
"Repositioned and replaced the sampling line, verified stable CO2 waveform and respiration rate detection, and returned the monitor to service."

Helpful Details to Include (If Known)

Final Thought

No breath and apnea alarms must be treated as patient safety events first. Clinical Engineering should verify the patient is safe, then work logically through the external causes: sampling line, patient connection, accessory condition, setup, and alarm settings. If the monitor still cannot detect breaths with known-good accessories, remove it from service and document the troubleshooting clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide