On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Addresses recurring sampling blockage caused by condensation, secretions, humidification, accessory placement, tubing routing, or unsuitable disposable components.
Step-by-Step Troubleshooting
1. Maintain Patient Monitoring
Move the patient to another verified capnography device or provide an approved alternate respiratory-monitoring method before disconnecting a blocked sampling line.
Do not allow the patient to remain dependent on a monitor with repeated waveform or sample loss.
Expected outcome: Respiratory monitoring continues while the affected setup is evaluated.
2. Confirm the Pattern of Blockage
Determine how often the blockage occurs, whether it follows suctioning, nebulizer treatment, humidification, patient repositioning, coughing, or transport, and whether the same accessory type is repeatedly involved.
Record any displayed blockage or FilterLine messages.
Expected outcome: The recurring condition is clearly associated with a specific setup, treatment, or accessory pattern when possible.
3. Remove and Replace the Contaminated FilterLine
Disconnect the blocked or wet FilterLine and inspect it externally for visible condensation, secretions, or contamination. Dispose of single-use accessories according to facility policy.
Do not flush, blow through, dry, or attempt to reuse the contaminated line.
Expected outcome: A new compatible FilterLine restores sample flow and waveform display. If the condition does not recur, troubleshooting can stop after observation.
4. Inspect Patient-Side Positioning
Verify that the cannula or airway adapter is correctly positioned and not placed where pooled secretions can enter the sampling inlet.
Check for loose airway fittings, dependent tubing loops, or an orientation that allows liquid to drain toward the sampling line.
Expected outcome: The sampling point is positioned to capture exhaled gas while minimizing liquid entry.
5. Review Tubing Routing
Route the sampling tubing without low loops, compression, or contact with wet surfaces. Ensure that bedding, patient movement, equipment mounts, and transport hardware do not alter the line position.
Expected outcome: The line remains supported, unobstructed, and less likely to collect condensation or secretions. If repeated blockage stops, troubleshooting can stop.
6. Verify Accessory Compatibility and Application
Confirm that the installed FilterLine and patient interface are compatible with the monitor and appropriate for the intended airway and oxygen-delivery setup.
Do not substitute unapproved tubing, adapters, or improvised connectors.
Expected outcome: The accessory selection is appropriate for the clinical application and remains recognized by the monitor.
7. Evaluate Humidification and Aerosol Sources
Determine whether heated humidification, nebulized medication, airway secretions, or high-condensation conditions are contributing to repeated contamination.
Coordinate with respiratory therapy or the clinical team to correct the patient setup without changing prescribed therapy.
Expected outcome: Sources of excessive moisture are identified and managed while maintaining clinical treatment.
8. Inspect the Monitor’s FilterLine Port
With the line disconnected, inspect the monitor port externally for visible moisture or residue. If contamination appears to have entered the monitor, remove it from service.
Do not insert swabs, tools, compressed air, or liquid cleaner into the port unless specifically authorized by manufacturer instructions.
Expected outcome: The port is visibly clean and dry. Any suspected internal contamination results in service escalation.
9. Observe With a New Line Under Clinical Conditions
After correcting positioning and routing, install a new FilterLine and observe waveform and sample-flow stability during typical patient care activities.
Expected outcome: The sampling path remains open and produces a stable waveform without repeated blockage. If so, troubleshooting can stop.
10. Escalate Recurrent Blockage With Correct Setup
If blockage continues despite a correct patient setup, compatible new accessories, and appropriate routing, remove the monitor from service and label it Out of Service.
Bench evaluation is required to assess the FilterLine receptacle, internal sampling pathway, flow detection, and pump performance.
Expected outcome: A monitor that may be retaining moisture or developing inadequate flow is withheld from clinical use.
If the Problem Persists
Common external causes have been addressed. Remaining possibilities include internal contamination, reduced pump performance, internal pneumatic restriction, damaged connector components, or a clinical application requiring manufacturer or respiratory-therapy review.
The device should be removed from service, labeled Out of Service, and evaluated using approved equipment and manufacturer documentation. Internal cleaning, pneumatic repair, or component replacement must be performed only by qualified personnel.
After repair, verify stable sampling over an extended test, correct waveform display, EtCO2 response, alarms, and FilterLine recognition before return to service.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Replace a contaminated sampling line promptly; a partially blocked line may display delayed or misleading EtCO2 values before complete failure.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Respiratory Therapy reported repeated sampling blockage and waveform loss during humidified airway monitoring."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found a dependent loop in the FilterLine that was collecting condensation and allowing moisture to enter the sampling path."
Resolution
What action was taken.
Example:
"The FilterLine was replaced and rerouted without a dependent loop, and stable waveform and sample flow were verified during continued observation."
Helpful Details to Include (If Known)
- Exact blockage message
- Frequency and timing of recurrence
- Airway and oxygen-delivery setup
- Humidification or nebulizer use
- Visible moisture or secretions
- FilterLine type
- Tubing orientation and routing
- Monitor port condition
- Results after accessory replacement
- Final device status
Final Thought
Protect the patient, replace contaminated disposables, and correct positioning and routing before suspecting an internal failure. Persistent moisture entry requires removal from service, controlled evaluation, and clear documentation of the conditions that caused the problem.
That is successful troubleshooting.