Medtronic McGRATH MAC

Power Button, Screen Hinge, or Handle Damage

On this page

Asset Type

Video Laryngoscope

Manufacturer

Medtronic

Model

McGRATH MAC

What This Guide Helps With

Evaluating reported mechanical damage affecting the power button, screen hinge, handle integrity, or safe reliable operation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Remove the McGRATH MAC from clinical use if the handle, power button, or screen hinge appears damaged, loose, cracked, unstable, or unreliable.

If the device is needed for an airway procedure, provide clinical staff with another verified functional video laryngoscope before troubleshooting.

Expected outcome: The affected device is evaluated away from patient care, with no airway procedure depending on damaged equipment.

Why it matters: A damaged video laryngoscope may fail during airway visualization or become unsafe to manipulate during intubation.

2. Confirm the Reported Damage

Ask staff what occurred before the concern was noticed, such as a drop, impact, difficulty powering on, loose display, or visible handle damage.

Inspect the exterior for:

Do not continue functional testing if sharp edges, exposed internal components, fluid intrusion, or major structural damage is present.

Expected outcome: The damage is clearly characterized and severe damage is immediately identified for removal from service.

3. Inspect the Blade Interface and Camera Area

With no blade installed, visually inspect the distal blade mounting area and camera/light window for cracks, deformation, residue, or damage related to the reported impact.

Install a new compatible disposable blade only if the handle remains structurally intact and safe to test. Confirm the blade attaches normally without unusual force, looseness, or misalignment.

Expected outcome: The blade attaches securely and the camera/light area has no visible damage.

If the blade cannot be attached normally or the interface is damaged, stop troubleshooting and remove the device from service.

4. Check the Battery Connection and External Power Condition

Inspect the installed compatible battery for secure seating, damage, contamination, swelling, or cracked housing. Reseat the battery as appropriate, or install a known-good compatible battery if battery condition is questionable.

The McGRATH MAC uses a battery-powered handle and provides battery status information during operation.

Expected outcome: A known-good battery is securely installed and the device is ready for a controlled power-on test.

If the battery compartment, battery connection area, or surrounding handle housing is damaged, stop and remove the device from service.

5. Evaluate Power Button Operation

Press the power button normally once while observing the device response. Do not repeatedly force or manipulate a damaged button.

Check whether:

The McGRATH MAC is designed with single-button functionality; unreliable actuation may prevent dependable clinical operation.

Expected outcome: The device powers on consistently with normal button actuation.

If the button sticks, requires excessive pressure, fails intermittently, or causes unstable operation, remove the device from service.

6. Evaluate the Screen and Hinge Without Forcing Movement

With the device powered on and unoccupied by a patient, gently move the screen only through its normal positioning range. Do not force a resistant or damaged hinge.

Inspect for:

The McGRATH MAC handle includes a reinforced screen hinge; visible looseness, breakage, or loss of stable positioning warrants repair evaluation.

Expected outcome: The screen adjusts normally, stays in position, and maintains a stable image.

If hinge damage or unstable screen positioning is found, stop troubleshooting and remove the device from service.

7. Verify Image and Illumination After External Inspection

If the device remains mechanically intact and safe to test, use a new compatible disposable blade and confirm:

Do not use the device clinically merely because an image is present if the handle, hinge, or power button remains damaged or unreliable.

Expected outcome: The device provides a stable image with no evidence that the reported damage affects operation.

8. Determine Final Device Status

Return the device to service only when all of the following are true:

Expected outcome: The final status is clear: safely returned to service after successful evaluation, or removed from service for repair.

If the Problem Persists

When visible damage, unreliable power-button operation, hinge instability, handle cracking, or image interruption remains after basic external checks, common easily correctable causes have been ruled out. The issue is likely a mechanical or internal device failure.

The McGRATH MAC should be:

Do not attempt internal handle, hinge, button, or display repairs as part of basic troubleshooting. Knowing when to stop and escalate is proper Clinical Engineering troubleshooting.

Clinical Use Tip

Do not troubleshoot a damaged video laryngoscope during an active airway procedure. Ensure another functional airway visualization device is immediately available before evaluating the affected unit.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia reported that the McGRATH MAC display hinge was loose and the screen would not remain positioned during setup for intubation."

Cause

What was observed during troubleshooting.

Example:
"Inspection found excessive hinge movement with unstable screen positioning; power-on image remained present, but the display could not be reliably positioned for clinical use."

Resolution

What action was taken.

Example:
"Removed the McGRATH MAC from service, applied an Out of Service label, provided the department with a replacement unit, and routed the damaged handle for repair evaluation."

Helpful Details to Include (If Known)

Final Thought

A video laryngoscope with mechanical damage must be evaluated as an airway safety concern, not simply a cosmetic issue. Begin with patient safety, confirm external damage and function logically, and remove the unit from service whenever dependable use cannot be verified. Clear CCR documentation helps protect patients, staff, and the repair process.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide