On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide assists Clinical Engineering in troubleshooting poor signal quality or frequent artifact alarms on the Nellcor PM100N. These issues are commonly caused by external factors such as sensor placement, patient movement, cable condition, or environmental interference rather than internal device failure.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Confirm the patient is on a functioning monitoring device before troubleshooting.
- Do not troubleshoot while the device is actively in use on a patient.
Verify Sensor Connection
- Ensure the SpO₂ sensor is fully seated in the monitor.
- Check for loose or intermittent connections.
- A poor connection can cause signal dropouts and artifact alarms.
Inspect the Sensor Condition
- Check for damaged, worn, or contaminated sensors.
- Look for cracked housings, frayed cables, or dirty optical surfaces.
- Replace the sensor if any defects are found.
Confirm Proper Sensor Type
- Verify that the correct Nellcor-compatible sensor is being used.
- Incompatible or non-approved sensors may produce unreliable readings.
Check Sensor Placement
- Ensure the sensor is applied correctly (finger, toe, or ear depending on type).
- Confirm proper alignment of emitter and detector.
- Poor placement is one of the most common causes of artifact alarms.
Assess Patient Movement
- Determine if the patient is moving excessively, shivering, or restless.
- Motion artifact is a frequent cause of poor signal quality.
- Test the unit on a stationary subject or simulator if possible.
Evaluate Perfusion at Site
- Check for low perfusion conditions such as cold extremities or poor circulation.
- Try a different site if perfusion is questionable.
- Low perfusion reduces signal strength and stability.
Inspect Extension Cables (if used)
- Remove any extension cable and connect the sensor directly to the monitor.
- Swap cables if available to rule out cable failure.
Check for Ambient Light Interference
- Ensure the sensor site is not exposed to bright light sources such as surgical lights or sunlight.
- Cover the sensor if needed to reduce interference.
Review Alarm Settings
- Verify alarm thresholds and sensitivity settings are appropriate.
- Overly sensitive settings may trigger frequent artifact alarms.
Power Cycle the Device
- Turn the monitor off and back on.
- This can clear temporary processing or signal acquisition issues.
Test with Known Good Sensor
- Use a verified working sensor to determine if the issue follows the sensor or the device.
- This helps isolate accessory versus device-related problems.
If the Problem Persists
If all external factors have been ruled out and the issue continues, the problem is likely internal such as signal processing circuitry or sensor interface failure.
- Remove the device from service
- Label it Out of Service
- Send for repair or perform bench evaluation
Recognizing when external troubleshooting has been exhausted is an important part of proper Clinical Engineering practice.
Clinical Use Tip
Do not troubleshoot signal quality issues on an active patient. Move the patient to a properly functioning pulse oximeter before beginning troubleshooting to ensure continuous and safe monitoring.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
“Unit displaying frequent artifact alarms and unstable SpO₂ readings during use.”
Cause
What was observed during troubleshooting.
Example:
“Worn SpO₂ sensor cable with intermittent connection causing signal dropouts.”
Resolution
What action was taken.
Example:
“Replaced defective sensor with known good unit. Verified stable readings and no further alarms.”
Helpful Details to Include (If Known)
- Sensor condition and type verified
- Sensor placement checked and adjusted
- Alternate site tested
- Extension cable removed or swapped
- Known good sensor tested
- Alarm behavior observed
- Ambient light conditions assessed
- Final device status documented
Final Thought
Most pulse oximeter signal issues are caused by external factors such as sensor condition, placement, or patient-related variables. A structured approach that rules out these common causes first ensures efficient troubleshooting while protecting patient safety. Proper escalation and clear documentation complete the process.
That is successful troubleshooting.