On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots abnormal electrosurgical effect associated with accessory condition, connection, technique-related setup, operating mode, power selection, or generator performance.
Step-by-Step Troubleshooting
1. Protect the Patient and Stop Unreliable Electrosurgical Use
Unexpected tissue effect, excessive arcing, or poor coagulation can create burn or surgical injury risk.
Do not continue troubleshooting on a patient. Provide a verified alternate ESU or approved surgical method and remove the affected setup from active use.
Expected outcome: The patient is no longer exposed to an unreliable electrosurgical system.
2. Clarify the Reported Tissue-Effect Problem
Determine what staff observed:
- Poor or absent coagulation
- Excessive sparking or arcing
- More aggressive tissue effect than expected
- Intermittent cutting or coagulation
- Abnormal effect with one accessory only
- Abnormal effect in one mode only
- A change after replacing an electrode or cable
Document the accessory, mode, and general settings in use without changing them prematurely.
Expected outcome: The reported condition is defined well enough to reproduce safely on test equipment.
3. Inspect the Active Electrode and Accessory
Inspect the active electrode, pencil, bipolar forceps, and associated cable as applicable.
Look for:
- Damaged insulation
- Loose electrode seating
- Carbon buildup or contamination
- Bent components
- Damaged cable
- Exposed conductor
- Loose connector
- Noncompatible accessory connection
Replace obviously damaged accessories before further testing.
Expected outcome: Only intact, compatible accessories are used during evaluation.
If replacing a damaged or contaminated accessory restores normal analyzer performance, troubleshooting may stop after final verification.
4. Verify All External Connections
Confirm active-accessory connections are secure.
For monopolar operation, inspect the patient return-electrode cable and connector as part of the test setup. Do not attempt clinical testing on a patient.
Check for looseness or intermittency at external receptacles.
Expected outcome: All required accessory connections are stable and fully seated.
If reseating a loose connection eliminates intermittent output during bench testing, continue to final verification.
5. Verify the Selected Mode and Power Setting
Confirm the operating mode and power setting reported by staff are appropriate for the intended clinical setup and were not inadvertently changed.
Clinical Engineering should verify control operation and displayed selections rather than prescribe surgical settings.
Compare actual control response with the displayed mode and setting.
Expected outcome: Controls select and display the intended mode and power consistently.
If an inadvertently changed control setting caused the complaint and the device tests normally afterward, troubleshooting can stop after documentation and verification.
6. Test With Known-Good Accessories
Replace the reported active accessory and applicable cables with known-good compatible components.
Repeat controlled activation into an approved electrosurgical analyzer or test load.
Expected outcome: Output is stable and repeatable with known-good accessories.
If the abnormal effect disappears, remove the original accessory from use and complete return-to-service testing.
7. Evaluate Output Stability With an Electrosurgical Analyzer
Test the applicable modes using approved test equipment and your organization’s authorized procedure.
Observe for:
- Stable activation
- Repeatable measured output
- Unexpected dropout
- Abnormal fluctuation
- Activation when not commanded
- Unexpected audible or visual behavior
Do not infer tissue performance from open-air arcing alone.
Expected outcome: Generator output is stable and consistent under controlled testing.
If output is unstable or inconsistent with expected performance, remove the generator from service.
8. Check for Environmental and Setup Contributors
Inspect the immediate operating setup for external conditions that can contribute to apparent arcing or inconsistent effect, including damaged accessory insulation, wet connections, poorly seated electrodes, or accessory cable stress.
Do not modify surgical technique or direct clinicians on tissue application.
Expected outcome: No external equipment or environmental condition is identified that would explain the complaint.
9. Perform Final Functional Verification
Following correction:
- Verify mode selection.
- Verify power-control operation.
- Verify accessory connections.
- Verify stable output using approved test equipment.
- Confirm activation indications and alarms.
- Perform required electrical safety and return-to-service checks.
Expected outcome: The ESU performs consistently and safely through the required test range.
If all required tests pass, troubleshooting is complete.
If the Problem Persists
If accessory damage, loose connections, incorrect control selection, and other external causes have been ruled out and measured output remains abnormal or unstable, further clinical use is inappropriate.
Potential remaining categories include internal generator regulation, sensing, control, or output-stage problems that require service-level evaluation.
The ESU should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate manufacturer documentation and approved electrosurgical test equipment
- Repaired or configured only by qualified personnel
Do not attempt board-level troubleshooting during routine field evaluation.
Return the ESU to service only after the underlying fault is corrected and output, controls, alarms, and safety functions pass required testing.
Clinical Use Tip
Unexpected tissue effect should be treated as an equipment-safety concern until the entire ESU-accessory path has been verified away from the patient.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported excessive arcing and inconsistent coagulation while using the Valleylab Force FX."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found carbon buildup and damage on the active electrode while generator output remained stable with a known-good accessory."
Resolution
What action was taken.
Example:
"Replaced the defective active accessory, verified stable Cut and Coag output with an electrosurgical analyzer, and returned the ESU to service after testing."
Helpful Details to Include (If Known)
- Tissue-effect complaint described by staff
- Operating mode involved
- Displayed power setting
- Active accessory condition
- Return-electrode cable condition when applicable
- Known-good substitution results
- Presence of visible arcing or intermittent activation
- Analyzer results
- Control operation
- Final device status
Final Thought
Abnormal tissue effect does not automatically indicate generator failure. Protect the patient, inspect the complete accessory path, verify controls and setup, measure output with approved equipment, and escalate when performance remains abnormal.
That is successful troubleshooting.