Medtronic Valleylab FT10

Error: E310 Touchscreen Error

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

Medtronic

Model

Valleylab FT10

What This Guide Helps With

This guide helps troubleshoot Error E310 Touchscreen Error on the Medtronic Valleylab FT10 electrosurgical unit. This condition may involve touchscreen input faults, unintended selections, unreliable control response, or unsafe changes during ESU setup.

Because the touchscreen is used for setup, mode selection, and power adjustments, any touchscreen reliability concern should be treated seriously before the device is used clinically.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Inspect the Touchscreen Surface

Check for Continuous or Ghost Touch Input

Check the Chassis and Display Area

Verify External Conditions

Power Cycle After Inspection and Cleaning

Determine Whether the Issue Is External or Device-Related

If the Problem Persists

If Error E310 reappears after the touchscreen is cleaned, dried, inspected, and isolated from external pressure, external causes have been ruled out.

Remove the Valleylab FT10 from service.

Label the unit Out of Service.

Send the device for repair or bench evaluation.

Persistent E310 conditions may indicate a touchscreen/display assembly issue, Display PCBA fault, chassis-related pressure problem, or other internal failure requiring service-level repair.

Clinical Use Tip

Never troubleshoot an ESU touchscreen input problem on an active patient. Move the patient or procedure to a backup device first, then troubleshoot the Valleylab FT10 in a safe service area.

An ESU with unresolved touchscreen input errors should not remain in surgical use because touchscreen faults can affect setup confidence, mode selection, and power setting reliability.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Valleylab FT10 displayed Error E310 Touchscreen Error. Staff reported touchscreen concern, possible unintentional screen input, and medium priority alarm condition."

Cause

What was observed during troubleshooting.

Example:
"Touchscreen inspected for contamination, residue, damage, and external pressure. Error persisted after cleaning, inspection, drying time, and power cycle. Likely Display PCBA or chassis-related touchscreen fault."

Resolution

What action was taken.

Example:
"Removed Valleylab FT10 from service, labeled Out of Service, and sent for repair or bench evaluation due to recurring E310 touchscreen error."

Helpful Details to Include (If Known)

Final Thought

Error E310 should be treated as a touchscreen reliability issue, not just a nuisance message. Start with patient safety, check for contamination or pressure first, then escalate when the error persists. Clear documentation protects the patient, the surgical team, and the repair process.

That is successful troubleshooting.